Image processing parameters (only for ROW)
The default values of image processing parameters will be set automatically by the SW
installation. According to the reporting physician´s image expression, the parameters
Factor Make Window Width Wider
and
Window Center Shift
can be adjusted to obtain
the best results. For more descriptions, see the table Description of Image Processing
Parameters.
For these settings, select
Local service >> Acquisition system >> General settings
>> Image processing
; see
.The default processing parameter groups
cannot be overwritten. For a customer's personal settings, a new group, e.g. Standard,
has to be created.
Fig. 59:
Image processing parameters
Description of Image Processing Parameters
Parameter
Default
Value
Description
Changeable
Additional Edge
Enhancement Gain
-0.8
Gain value for additional noise fil-
tering
Do not change
Additional Edge
Enhancement Kernel
3
Kernel value for additional noise fil-
tering
Do not change