84
Image Quality
MAMMOMAT Novation DR
SPB7-250.815.06.02.02
Siemens AG
12.05
CS SD 24
Page 84 of 116
Medical Solutions
Amplification Factor
2.52
Fix value for amplification, only
used when mode = OFF
Do not change
Amplification Mode
ON
See above
Do not change
Set Direct Radiation
Black
ON
Takes care that direct radiated area
is set to black
Do not change
Window Mode
ON
ON: Complete area of breast is
taken into account for the calcula-
tion of the window/center values
OFF: only compressed area is
taken into account for the calcula-
tion of the window/center values
Do not change
Edge Enhancement
Gain
0.3
Gain value for enhancement of the
local contrast
Do not change
Edge Enhancement
Kernel
15
Do not change
Edge Enhancement
LUT index
0
Currently not used; future plan to
have different edge enhancement
based on mean gray value in image
Do not change
LUT index
18
Separate LUTs to calculate the dis-
played values are delivered
Do not change
Harmonization Ampli-
fication Gain
0.7
Gain value for enhancement of the
global contrast
Do not change
Harmonization Kernel
Size
127
Gain value for enhancement of the
global contrast
Do not change
Factor make window
width wider
1.6
Defines range of gray values to be
displayed
Decrease if
image should
have more con-
trast; increase if
image should
have less con-
trast;
3 predefined
settings exist:
Normal, Low,
High
Parameter
Default
Value
Description
Changeable