62
Checking dose settings
MAMMOMAT Novation DR
SPB7-250.815.06.02.02
Siemens AG
12.05
CS SD 24
Page 62 of 116
Medical Solutions
4. Click
Modify settings
.
Fig. 40: BRICK_Modify settings
5. To adjust the
-
H
value,
go to
Modify low dose scaling
(The default values are 80, 70, 45) and enter the new
values, separated by a comma, e.g.
80,70,45
in the
Low Dose Scaling Value
field and
click
Change Low Dose Scaling
to save it.
-
D
value,
go to
Modify high dose scaling
(The default values are 100, 80, 60) and enter the new
values, separated by a comma, e.g.
100,80,60
in the
High Dose Scaling Value
field
and click
Change High Dose Scaling
to save it.
6. Proceed from where you were originally referenced.