CS SD 24
Checking dose settings
61
Page 61 of 116
Medical Solutions
P
5. Record the measured values in the "
Start-up Protocol
" on
.
Adjust the H and D Settings
To adjust the
enter the new values, separated by a comma, e.g.
100,80,60.
The values represent the
factor in % for (Mo/Mo, Mo/Rh, W/Rh).
Three settings, e.g.
try
(short-term, is valid for one measurement only),
save
(long-term,
valid also after reboot) and
load
(effective, valid until the system is rebooted) can be used
in the drop-down menu above the values.
To change the values permanently, enter the correct values and select
save
, then use
load
to load them before you start another exposure.
-->
Calculation of the new Mo/Mo dose scaling value:
-->
Calculation of the new Mo/Rh, W/Rh dose scaling values:
-->
Proceed as follows:
1. Call up the service software and select:
Acquisition System -> BRICK Configuration
The BRICK service software opens another window.
2. Select
Brick -> AEC Configuration
.
3. The BRICK software switches to a higher user level, thus a Username and Password
have to be entered.
User name:
root
Password:
******
H
value, go to
Modify low dose scaling
(80, 70, 45) and
D
value, go to
Modify high dose scaling
(100, 80, 60) and
NewScalingValue = OldScalingValue * DoseValue
Nominal
/ DoseValue
Measured
DoseValue
Nominal
for H = 130
µ
Gy
DoseValue
Nominal
for D = 190
µ
Gy
NewScalingValue = OldScalingValue * PixelValue
Mo/Mo
/ PixelValue
Measured