Siemens AG
SPB7-250.815.06.02.02
MAMMOMAT Novation DR
12.05
CS SD 24
Checking dose settings
57
Page 57 of 116
Medical Solutions
Checking OPDOSE
0
The OPDOSE values are checked in this chapter.
In AUTO mode the system suggests an appropriate program (kV, Anode/Filter combina-
tion) corresponding to a specific breast thickness. The new, suggested program is indi-
cated by a blinking LED, whereas the previously used program is shown by a
continuously lighted LED.
Please see
the Instructions of Use
(SPB7-250.620.01...) for the correct OPDOSE val-
ues.
Changes can be made with the MAMMOMAT stand service software:
Configuration -> Miscellaneous -> Auto limits
.
A special setting can be stored for each wing, thus the measurement has to be performed
for the
detector wing
and the
OPDIMA wing
.
Proceed as follows for the "detector wing":
1. Select the
detector wing. dr
is then displayed on the control console.
2. Mount a 18x24 cm compression plate.
3. Select a thickness of calibration plexi (part no. 65 61 240) that is within the thickness in-
terval of the program to be tested.
4. Select
AUTO
on the control panel.
5. Place the calibration plexi on the object table and compress to 6 kg or more.
6. Check that the correct program is blinking on the control panel. Select the program by
pressing the corresponding program button on the control panel.
7. Repeat the procedure described in
steps 3-6
for the other programs used.
P
8. Record the measured values in the "
Installation Protocol
" (SPB7-250.813.01...).
Proceed as follows for the “OPDIMA wing”:
1. Select the
OPDIMA wing
. As a result,
OP
is displayed on the control console.
2. Mount the
spot 1.0
compression plate.
3. Select a thickness of calibration plexi (part no. 65 61 240) that is within the thickness in-
terval of the program to be tested.
4. Select
AUTO
on the control panel.
5. Place the calibration plexi on the object table and compress to 6 kg or more.
6. Check that the correct program is blinking on the control panel. Select the program by
pressing the corresponding program button on the control panel.
7. Repeat the procedure described in
steps 3-6
for the other programs used.
P
8. Record the measured values in the "Start-up Protocol" on
.