Medical Solutions
Delay times between two exposures
0
The delay times listed below must be followed to prevent the tube from overheating.
Tab. 4
Delay times between two exposures
Exposure mAs value
Delay time between two exposures (seconds)
max. 100
min. 15
max. 200
min. 30
max. 300
min. 45
max. 400
min. 60
max. 500
min. 75