
9410 series
Inputs / outputs
2. Open the Setup Assistant for your meter.
3. Navigate to
Expansion Module > Option I/O Modules
and select the module tab
that corresponds to the option module with the analog input you want to configure.
4. Select the analog input channel and click
Edit
. The setup screen for that analog
input is displayed.
5. Configure the analog input by selecting the parameter and clicking
Edit
.
Parameter
Value/Range
Description
Zero scale
0 to ±10
9
The minimum source value that matches the minimum
analog input signal.
Full scale
0 to ±10
9
The maximum source value that matches the maximum
analog input signal.
Port
–
The physical analog input port connection.
Mode
Voltage, current
Determines whether the analog input is monitoring voltage
or current.
Analog outputs
Analog output applications
You can use an analog output to send a signal to an analog-controlled valve to modify
a water pipe’s valve position in order to change the flow rate.
For analog output operation, your meter takes an input value and scales it to the
appropriate signal value to send out the physical analog output port.
Your meter’s analog I/O option modules can output either low voltage or current using
standard analog transducers such as 4 - 20 mA current transducers.
Related Topics
•
Specifications
Analog output behavior
Your meter’s analog output, once configured, outputs voltage or current signals. If your
meter’s analog output source becomes unavailable (N/A) the analog output level
should go to zero (0 V / 0 mA) under most conditions.
When your meter is powered off or configured to send voltage, the meter registers as a
high impedance.
The electrical signal on the analog output is DC; make sure proper polarity is observed
when wiring external devices to the analog output ports.
An unexpected change of state of the analog outputs can result when the supply
power to the meter is interrupted, during an upgrade of the meter’s firmware or
framework or option module firmware or a during an option module reset.
WARNING
UNINTENDED OPERATION
Failure to follow these instructions can result in death, serious injury, or
equipment damage.
Do not use this device for critical control or protection applications where human or
equipment safety relies on the operation of the control circuit.
116
7EN05-0336-01