Functions
7.9 Analog Output Trim
FUP1010 IP67 Portable
Operating Instructions, 03/2012, A5E02951522-02
97
3.
To enable numeric entry, press the <Right Arrow>. Type a numeric value corresponding
to a 4mA input signal.
4.
To store the data press <ENTER>. This moves the cursor to [20 mA].
5.
To enable numeric entry, press the <Right Arrow>. Type the numeric value
corresponding to a 20mA input signal.
6.
To store the data, press <ENTER>.
7.9
Analog Output Trim
Introduction
Analog Out Trim function allows you to fine-tune the flow meter’s analog voltage and current
outputs using a multi-meter connected to the output under test. In addition, you can use a
frequency counter to fine-tune the flow meter’s pulse rate output.
Note
The current, voltage, and Pgen trimming will be limited by the 12-bit resolution of the flow
meter’s D/A Convertor (DAC).
1.
From the [Meter Facilities] menu, scroll to the [Analog OutTrim] menu.
2.
Press the <Right Arrow> to access the option list.
0HWHU)DFLOLWLHV
6LHPHQV
'XDO3DWK>@3DWK
$QDORJ2XWSXW7ULP
3UHIHUUHG8QLWV
(QJOLVK
7DEOH6HWXSV
/RJJHU&RQWURO
0HPRU\&RQWURO
$QDORJ2XW7ULP
57'&DOLEUDWH
&ORFN6HW
566HWXS
1RQH>@
%DFNOLJKW
2Q
6\VWHP,QIR
Table 7- 8
Analog Out Trim Menu Structure
Trim Io1/Io2
Operate
Trim@4mA Indicated mA=x.xx
Analog Out Trim
Trim Vo1/Vo2
Operate
Trim@2V Indicated V=x.xx
Trim Pgen1 / Pgen2
Operate
Trim@1 Hz Indicated Hz=xxxx
Summary of Contents for 7ME3510
Page 2: ......
Page 10: ...Table of contents FUP1010 IP67 Portable 8 Operating Instructions 03 2012 A5E02951522 02 ...
Page 159: ...1010WP 7 21614 C ...
Page 160: ...1010WP 7 21614 C ...
Page 161: ...1010WDP 7 21614 C ...
Page 162: ...1010WDP 7 21614 C ...
Page 163: ......
Page 164: ......
Page 165: ......
Page 167: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Page 168: ......
Page 169: ...21614 C ...
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ...21614 C ...
Page 175: ...21614 C ...
Page 176: ...21614 C ...
Page 177: ...21614 C ...
Page 178: ...21614 C ...
Page 179: ...21614 C ...
Page 180: ...21614 C ...
Page 181: ......
Page 182: ......
Page 183: ...21614 C ...
Page 184: ...21614 C ...
Page 185: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Page 186: ...REFLECT MODE DIRECT MODE 21614 C ...
Page 187: ...1015BC 1 8 21614 C ...
Page 188: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Page 189: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Page 190: ......
Page 191: ......
Page 192: ......
Page 193: ......
Page 195: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Page 196: ......
Page 197: ......
Page 208: ...Glossary FUP1010 IP67 Portable 162 Operating Instructions 03 2012 A5E02951522 02 ...
Page 212: ...Index FUP1010 IP67 Portable 166 Operating Instructions 03 2012 A5E02951522 02 ...
Page 213: ......