Troubleshooting
10.3 Test Facilities Graph Screen
FUP1010 IP67 Portable
Operating Instructions, 03/2012, A5E02951522-02
129
[MinDamp #] Command
Pressing the <1> key will cause [MinDamp #] to appear on the command line at the lower
left-hand corner of the screen. The number listed to the right of the command code
represents the exponent in the meter exponential averaging routine, where the larger the
number the greater the digital averaging. Pressing the <+> key will increase the damping
value. Likewise, pressing the <-> key will decrease the damping value.
To exit this mode, press the <0> key on the keypad.
[MaxDamp #] Command
Pressing the <2> key will bring up the [MaxDamp #] command. The function of this
parameter is similar to the [MinDamp #] command described above; however, the two
parameters interact in the following manner. The MinDamp value must not exceed the
MaxDamp value; therefore increasing the MinDamp value above the previous MaxDamp
value will set both parameters to the same value. In most cases, it is preferred that both
damping parameters be set to the same value, however, in cases where rapid response to
changes in liquid sound velocity for flow rate is required, the two values may be set
differently. In this situation the meter will use the MaxDamp value when conditions are
stable, but then switch to a faster damping value (limited by MinDamp) when a significant
change in sound velocity or flow rate is perceived.
To exit this mode, press the <0> key on the keypad.
To access the Digital Damping Control using the Test Facilities Graph Screen, proceed as
follows:
Note
To use the Test Facilities Graph Screen you must have a working site.
Summary of Contents for 7ME3510
Page 2: ......
Page 10: ...Table of contents FUP1010 IP67 Portable 8 Operating Instructions 03 2012 A5E02951522 02 ...
Page 159: ...1010WP 7 21614 C ...
Page 160: ...1010WP 7 21614 C ...
Page 161: ...1010WDP 7 21614 C ...
Page 162: ...1010WDP 7 21614 C ...
Page 163: ......
Page 164: ......
Page 165: ......
Page 167: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Page 168: ......
Page 169: ...21614 C ...
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ...21614 C ...
Page 175: ...21614 C ...
Page 176: ...21614 C ...
Page 177: ...21614 C ...
Page 178: ...21614 C ...
Page 179: ...21614 C ...
Page 180: ...21614 C ...
Page 181: ......
Page 182: ......
Page 183: ...21614 C ...
Page 184: ...21614 C ...
Page 185: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Page 186: ...REFLECT MODE DIRECT MODE 21614 C ...
Page 187: ...1015BC 1 8 21614 C ...
Page 188: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Page 189: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Page 190: ......
Page 191: ......
Page 192: ......
Page 193: ......
Page 195: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Page 196: ......
Page 197: ......
Page 208: ...Glossary FUP1010 IP67 Portable 162 Operating Instructions 03 2012 A5E02951522 02 ...
Page 212: ...Index FUP1010 IP67 Portable 166 Operating Instructions 03 2012 A5E02951522 02 ...
Page 213: ......