Connecting
5.2 Sensor Wiring
FUP1010 IP67 Portable
30
Operating Instructions, 03/2012, A5E02951522-02
5.
Do not mount sensors from different ultrasonic flow meters on the same pipe. Also, do
not run the sensor cables in common bundles with cables from communication
equipment, other Siemens systems, or any type of ultrasonic equipment. You can run
these cables through a common conduit ONLY if they originate at the same flow meter.
6.
Never mount sensors under water, unless you order submersible units and you install
them in accordance with factory instructions.
7.
Never mount sensors on the top or bottom of a horizontal pipe. The best placement on a
horizontal pipe is either the ten o’clock and two o’clock position for Reflect Mode, or one
sensor at nine o’clock and one sensor at three o’clock for Direct Mode. Mounting on a
vertical pipe is recommended only if flow is in the upward direction. When mounting on a
vertical pipe flowing in a downward direction make sure there is sufficient back pressure
in the system to maintain a full pipe.
R
①
Dual Path, Reflect Mount
②
Dual Path, Direct Mount
Figure 5-4
Sensor Alignment (Horizontal Plane)
Preparing the Pipe
1.
Pick a mounting location with the longest straight run. You must have easy access to at
least one side of your pipe. The pipe at the mounting location must remain full, even at
zero flow.
2.
Decide on your mounting mode (Direct or Reflect). Always use Reflect Mode whenever
possible. You may only need to use Direct Mode if your pipe is plastic.
3.
After receiving the spacing dimensions from the Installation Menu, prepare the pipe
surface. De-grease the surface, if necessary, and remove any grit, corrosion, rust, loose
paint, etc. Use abrasive material provided to provide a clean contact surface for the
sensors.
Summary of Contents for 7ME3510
Page 2: ......
Page 10: ...Table of contents FUP1010 IP67 Portable 8 Operating Instructions 03 2012 A5E02951522 02 ...
Page 159: ...1010WP 7 21614 C ...
Page 160: ...1010WP 7 21614 C ...
Page 161: ...1010WDP 7 21614 C ...
Page 162: ...1010WDP 7 21614 C ...
Page 163: ......
Page 164: ......
Page 165: ......
Page 167: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Page 168: ......
Page 169: ...21614 C ...
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ...21614 C ...
Page 175: ...21614 C ...
Page 176: ...21614 C ...
Page 177: ...21614 C ...
Page 178: ...21614 C ...
Page 179: ...21614 C ...
Page 180: ...21614 C ...
Page 181: ......
Page 182: ......
Page 183: ...21614 C ...
Page 184: ...21614 C ...
Page 185: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Page 186: ...REFLECT MODE DIRECT MODE 21614 C ...
Page 187: ...1015BC 1 8 21614 C ...
Page 188: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Page 189: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Page 190: ......
Page 191: ......
Page 192: ......
Page 193: ......
Page 195: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Page 196: ......
Page 197: ......
Page 208: ...Glossary FUP1010 IP67 Portable 162 Operating Instructions 03 2012 A5E02951522 02 ...
Page 212: ...Index FUP1010 IP67 Portable 166 Operating Instructions 03 2012 A5E02951522 02 ...
Page 213: ......