Connecting
5.2 Sensor Wiring
FUP1010 IP67 Portable
40
Operating Instructions, 03/2012, A5E02951522-02
4.
Press the <Right Arrow> to [Sensor Model]. Press <Right Arrow> and scroll to select the
sensor model number on the sensor label.
5.
The drop down menu lists the following sensor selections:
–
1011 Universal
–
1011HP-T1 - Usable -40 to 120°C, recommended for Ø Temperature <40°C;
Standard.
–
1011HP-T2 - Usable -40 to 120°C, recommended for Ø Temperature >40°C - <80°C;
Named as high temperature.
–
1011HP-T3 - Usable -40 to 120°C, recommended for Ø Temperature >80°C <120°C;
special request.
–
991 Universal
Note
The meter will automatically recommend a sensor depending on the application data
that has been entered.
6.
For this example, select the sensor model that appears on the sensor label then press
<ENTER>.
,QVWDOO6HQVRU
6LHPHQV
&KDQQHO>@$%&
6FUROO/LVWDQGVHOHFWGHVLUHG0RGHO
6HQV37
6HQVRU6L]H
'+
6HQVRU0RXQW0RGH5HIOHFW
6SDFLQJ2IIVHW
1RPLQDO
1XPEHU,QGH[
6SDFLQJ0HWKRG 6SDFHU%DU%1
/WQ9DXHLQ
,QVWDOO&RPSOHWH1R
(PSW\3LSH6HW 07<PDWLF
=HUR)ORZ$GMXVW $FWXDO=HUR
①
Select based on type
②
Select based on size
③
After sensor is mounted select [Install].
7.
To select Sensor Size, press <Right Arrow>. Scroll to select the sensor size that matches
the size indicated on the sensor label. Press <ENTER>.
8.
At [Sensor Mount Mode], press the <Right Arrow>. Scroll to select [Reflect] or [Direct]
mount and then press <ENTER>.
9.
IMPORTANT: Record Spacing Method and Number Index. This data will be used to
mount the sensors.
10.
Sensors can now be mounted. Refer to mounting procedures and select the mounting
mode desired.
11.
After sensors are mounted scroll to [Install Complete] and select [Install].
Summary of Contents for 7ME3510
Page 2: ......
Page 10: ...Table of contents FUP1010 IP67 Portable 8 Operating Instructions 03 2012 A5E02951522 02 ...
Page 159: ...1010WP 7 21614 C ...
Page 160: ...1010WP 7 21614 C ...
Page 161: ...1010WDP 7 21614 C ...
Page 162: ...1010WDP 7 21614 C ...
Page 163: ......
Page 164: ......
Page 165: ......
Page 167: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Page 168: ......
Page 169: ...21614 C ...
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ...21614 C ...
Page 175: ...21614 C ...
Page 176: ...21614 C ...
Page 177: ...21614 C ...
Page 178: ...21614 C ...
Page 179: ...21614 C ...
Page 180: ...21614 C ...
Page 181: ......
Page 182: ......
Page 183: ...21614 C ...
Page 184: ...21614 C ...
Page 185: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Page 186: ...REFLECT MODE DIRECT MODE 21614 C ...
Page 187: ...1015BC 1 8 21614 C ...
Page 188: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Page 189: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Page 190: ......
Page 191: ......
Page 192: ......
Page 193: ......
Page 195: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Page 196: ......
Page 197: ......
Page 208: ...Glossary FUP1010 IP67 Portable 162 Operating Instructions 03 2012 A5E02951522 02 ...
Page 212: ...Index FUP1010 IP67 Portable 166 Operating Instructions 03 2012 A5E02951522 02 ...
Page 213: ......