Functions
7.9 Analog Output Trim
FUP1010 IP67 Portable
Operating Instructions, 03/2012, A5E02951522-02
99
To calculate a voltage output:
1.
Set up the multi-meter to read volts, then connect it to the supply and return terminals of
the voltage output under test.
2.
Move the highlight to the port to be tested by pressing <Up/Down Arrow>, press <Right
Arrow>, then press <Down Arrow> to move the cursor to [Trim @ 2V].
$QDORJ2XW7ULP
6LHPHQV
'XDO3DWK>@3DWK
7ULP9ROWRXWSXW
7ULP,R
2SHUDWH
7ULP,R
2SHUDWH
7ULP9R
!2SHUDWH
7ULP9R
7ULP#9
7ULP3JHQ
2SHUDWH
7ULP3JHQ
2SHUDWH
3.
Press <ENTER>. This triggers a 2.00 Volts pop-up window. The multi-meter should now
be reading 2.00 Volts.
4.
If the multi-meter reading does not match, use the numeric keys to type in the multi-meter
reading.
5.
Press <ENTER> to register setting. This adjusts the meter’s DAC (digital-to-analog
converter) so that a 2.00 Volts output corresponds with 2.00 Volts on the multi-meter.
6.
Re-check the multi-meter to make sure that it is now reading 2.00 Volts.
Pgen Output Trim
Note
Can be trimmed to within 1.25 Hz of nominal.
To calibrate a pulse rate output
1.
Connect a frequency counter to the signal and reference terminals of the pulse rate
output under test.
2.
Move the highlight to the port to be tested by pressing <Up/Down Arrow>, press <Right
Arrow> and then press <Down Arrow> to move the cursor to [Trim @ 1kHz].
3.
Press <ENTER>. This triggers a 1 kHz pop-up window. The frequency counter should
now be reading 1 kHz.
4.
If the frequency counter reading does not match, use the numeric keys to type in the
frequency counter reading.
Summary of Contents for 7ME3510
Page 2: ......
Page 10: ...Table of contents FUP1010 IP67 Portable 8 Operating Instructions 03 2012 A5E02951522 02 ...
Page 159: ...1010WP 7 21614 C ...
Page 160: ...1010WP 7 21614 C ...
Page 161: ...1010WDP 7 21614 C ...
Page 162: ...1010WDP 7 21614 C ...
Page 163: ......
Page 164: ......
Page 165: ......
Page 167: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Page 168: ......
Page 169: ...21614 C ...
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ...21614 C ...
Page 175: ...21614 C ...
Page 176: ...21614 C ...
Page 177: ...21614 C ...
Page 178: ...21614 C ...
Page 179: ...21614 C ...
Page 180: ...21614 C ...
Page 181: ......
Page 182: ......
Page 183: ...21614 C ...
Page 184: ...21614 C ...
Page 185: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Page 186: ...REFLECT MODE DIRECT MODE 21614 C ...
Page 187: ...1015BC 1 8 21614 C ...
Page 188: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Page 189: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Page 190: ......
Page 191: ......
Page 192: ......
Page 193: ......
Page 195: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Page 196: ......
Page 197: ......
Page 208: ...Glossary FUP1010 IP67 Portable 162 Operating Instructions 03 2012 A5E02951522 02 ...
Page 212: ...Index FUP1010 IP67 Portable 166 Operating Instructions 03 2012 A5E02951522 02 ...
Page 213: ......