Connecting
5.2 Sensor Wiring
FUP1010 IP67 Portable
Operating Instructions, 03/2012, A5E02951522-02
45
5.2.6
Direct Mount
Direct Mount - Installation with EZ Clamp and Spacing Guide
The combination of EZ Clamp and spacer guide is the recommended way to mount Direct
Mode sensors. This method ensures that sensors will align exactly 180° from each other and
remain spaced the proper distance apart.
For Direct Mode mounting, a spacer bar is used to establish the distance between sensors
and a spacing guide is used to locate the sensors at the nine o’clock and three o’clock
positions. Should the distance between sensors be beyond the span of a spacer bar, a
measuring tape can be used. The Mylar spacing guide comes in various lengths and widths
to accommodate most pipe sizes.
Spacing Guide Sizes
5.08cm x 66.04cm) (2" x 26")
5.08cm x 114.3cm (2" x 45")
10.16 cm x 205.7cm (4" x 81")
10.16cm x 393.7cm (4" x 155")
15.2cm x 497.8cm (6" x 196")
Figure 5-15 Mylar Spacing Guide
Installation Procedure
1.
Make a note of the Number Index displayed in the [Install Sensor] menu. Check to ensure
that you have a matched set of sensors. They both should have the same S/N number
but marked with either an "A" or "B" (e.g., 100A and 100B).
2.
Temporarily position one of the sensors on the pipe where you will be mounting it. Ensure
that this is a smooth area without any raised areas (seams, etc.) With a pencil or chalk,
mark a generous area around all sides of the sensor. Remove the sensor.
3.
Prepare the area you marked by degreasing the surface and removing any grit, corrosion,
rust, loose paint or surface irregularities with the abrasive pipe conditioning material
provided.
Summary of Contents for 7ME3510
Page 2: ......
Page 10: ...Table of contents FUP1010 IP67 Portable 8 Operating Instructions 03 2012 A5E02951522 02 ...
Page 159: ...1010WP 7 21614 C ...
Page 160: ...1010WP 7 21614 C ...
Page 161: ...1010WDP 7 21614 C ...
Page 162: ...1010WDP 7 21614 C ...
Page 163: ......
Page 164: ......
Page 165: ......
Page 167: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Page 168: ......
Page 169: ...21614 C ...
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ...21614 C ...
Page 175: ...21614 C ...
Page 176: ...21614 C ...
Page 177: ...21614 C ...
Page 178: ...21614 C ...
Page 179: ...21614 C ...
Page 180: ...21614 C ...
Page 181: ......
Page 182: ......
Page 183: ...21614 C ...
Page 184: ...21614 C ...
Page 185: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Page 186: ...REFLECT MODE DIRECT MODE 21614 C ...
Page 187: ...1015BC 1 8 21614 C ...
Page 188: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Page 189: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Page 190: ......
Page 191: ......
Page 192: ......
Page 193: ......
Page 195: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Page 196: ......
Page 197: ......
Page 208: ...Glossary FUP1010 IP67 Portable 162 Operating Instructions 03 2012 A5E02951522 02 ...
Page 212: ...Index FUP1010 IP67 Portable 166 Operating Instructions 03 2012 A5E02951522 02 ...
Page 213: ......