Commissioning
6.3 Empty Pipe Set
FUP1010 IP67 Portable
68
Operating Instructions, 03/2012, A5E02951522-02
Final Setup
1.
Press <MENU> key once to display the main menu.
6HOHFW0HWHU7\SH
0HWHU7\SH
!&KDQQHO)ORZ
0HWHU)DFLOLWLHV
'XDO3DWK)ORZ
/DQJXDJH
&K)ORZ
&K)ORZ
7KLFNQHVV*DXJH
6LHPHQV&KDQQHO>@&KDQQHO
2.
Press the <Right Arrow> twice.
3.
Scroll down to [Install Complete]. Press the <Right Arrow> and select [Install]. Press
<ENTER>. The flow meter will go through its drives.
4.
Observe the Measured Vs window and verify a correct sound velocity measurement (if
known).
5.
Press the <Down Arrow> to accept sound velocity value.
6.
The flow meter is ready to report flow. Press the <MENU> key again to display flow.
*0
%$7
)$8/7
63$&(
0$.83
6LHPHQV
&KDQQHO>@
8QLW
)/2:*$/0,1
727$/
*$/
9V
14.35
354.6597
14.27
100.0
0.0
31.74
9IR
6.3
Empty Pipe Set
The flow meter performs the MTYmatic routine automatically during its Initial Make-up to
establish a standard setting for the Empty Pipe alarm. This process is normally sufficient for
setting this parameter. The [Empty Pipe Set] option list allows you to re-invoke MTYmatic,
use an Actual MTY routine (if application conditions allow you to empty and refill the pipe) or
use the Set Empty routine to set the empty pipe threshold by direct numeric entry.
Summary of Contents for 7ME3510
Page 2: ......
Page 10: ...Table of contents FUP1010 IP67 Portable 8 Operating Instructions 03 2012 A5E02951522 02 ...
Page 159: ...1010WP 7 21614 C ...
Page 160: ...1010WP 7 21614 C ...
Page 161: ...1010WDP 7 21614 C ...
Page 162: ...1010WDP 7 21614 C ...
Page 163: ......
Page 164: ......
Page 165: ......
Page 167: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Page 168: ......
Page 169: ...21614 C ...
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ...21614 C ...
Page 175: ...21614 C ...
Page 176: ...21614 C ...
Page 177: ...21614 C ...
Page 178: ...21614 C ...
Page 179: ...21614 C ...
Page 180: ...21614 C ...
Page 181: ......
Page 182: ......
Page 183: ...21614 C ...
Page 184: ...21614 C ...
Page 185: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Page 186: ...REFLECT MODE DIRECT MODE 21614 C ...
Page 187: ...1015BC 1 8 21614 C ...
Page 188: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Page 189: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Page 190: ......
Page 191: ......
Page 192: ......
Page 193: ......
Page 195: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Page 196: ......
Page 197: ......
Page 208: ...Glossary FUP1010 IP67 Portable 162 Operating Instructions 03 2012 A5E02951522 02 ...
Page 212: ...Index FUP1010 IP67 Portable 166 Operating Instructions 03 2012 A5E02951522 02 ...
Page 213: ......