Troubleshooting
10.3 Test Facilities Graph Screen
FUP1010 IP67 Portable
128
Operating Instructions, 03/2012, A5E02951522-02
Time Base Control
The digitized receive signal can be moved either to the left or right on the screen by pressing
the <Left> or <Right> keypad arrows. The direction of the arrow actually represents the
direction in which the Receive "window" will move, thereby causing the receive signal to shift
in the opposite direction on the screen (e.g., Pressing the <Left Arrow> moves the signal to
the right).
The digitized receive signal can be expanded or contracted in the time domain by pressing
the <+> or <-> keys on the keypad. This allows you to see the entire contents of the receive
window, or zoom in to see greater detail. Pressing the <CLR> key once will automatically
center the receive signal on the screen. When expanding the Receive signal small vertical
"tick" marks will eventually appear. These marks represent the time at which the receive
signal is digitally sampled.
Correlated Plot
During conditions of flow, the actual transit time delta (difference) can be observed in the
displayed receive signal waveform when the [Correlated Plot] menu parameter is not
selected. To observe this time difference simply depress the <+> key (to see greater signal
detail) until the individual up and down receive signals are clearly discernible. To verify that
the flow meter signal processing algorithms are properly correlating the up and down stream
receive signals, select the [Correlated Plot] option from the display menu list.
Return to the graph screen and observe the relative position of the up and down waveforms.
In a properly correlated receive signal the two images should be nearly superimposed on top
of each other, even during high flow conditions. In the unlikely situation where the two
images appear to be offset by one or more receive cycles then the flow readings should be
considered questionable.
Command Modes
Although the flow meter signal processing algorithms are capable of accommodating a very
wide range of signal conditions, it may be desirable to override these default settings under
extremely difficult operating conditions. The following functions are available for this purpose.
Digital Damping Control: (Hot Key 1 and 2)
The meter permits user modification of the digital averaging used by the signal processing
routines. In general, the default damping values selected by the flow meter will provide
optimal performance over a wide range of transit time applications. However, in extreme
cases of unstable flow, pulsating flow, low signal levels or high electronic noise it may be
necessary to override these default settings to permit uninterrupted and reliable flow
measurement.
Test Facilities Graph Screen
The Graph Screen includes the capability to access a set of command codes, which enable
a user to override a number of default meter settings. The most important parameter is the
digital damping control, which can be accessed by pressing number <1> or <2> on the
keypad while in the Signal Graph Screen mode.
Summary of Contents for 7ME3510
Page 2: ......
Page 10: ...Table of contents FUP1010 IP67 Portable 8 Operating Instructions 03 2012 A5E02951522 02 ...
Page 159: ...1010WP 7 21614 C ...
Page 160: ...1010WP 7 21614 C ...
Page 161: ...1010WDP 7 21614 C ...
Page 162: ...1010WDP 7 21614 C ...
Page 163: ......
Page 164: ......
Page 165: ......
Page 167: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Page 168: ......
Page 169: ...21614 C ...
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ...21614 C ...
Page 175: ...21614 C ...
Page 176: ...21614 C ...
Page 177: ...21614 C ...
Page 178: ...21614 C ...
Page 179: ...21614 C ...
Page 180: ...21614 C ...
Page 181: ......
Page 182: ......
Page 183: ...21614 C ...
Page 184: ...21614 C ...
Page 185: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Page 186: ...REFLECT MODE DIRECT MODE 21614 C ...
Page 187: ...1015BC 1 8 21614 C ...
Page 188: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Page 189: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Page 190: ......
Page 191: ......
Page 192: ......
Page 193: ......
Page 195: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Page 196: ......
Page 197: ......
Page 208: ...Glossary FUP1010 IP67 Portable 162 Operating Instructions 03 2012 A5E02951522 02 ...
Page 212: ...Index FUP1010 IP67 Portable 166 Operating Instructions 03 2012 A5E02951522 02 ...
Page 213: ......