
Troubleshooting
10.2 F4 Reset Procedure
FUP1010 IP67 Portable
Operating Instructions, 03/2012, A5E02951522-02
125
Note
If you receive a Detection Fault message, it is strongly recommended that the Technical
Service Department (
http://www.automation.siemens.com/partner
) be contacted.
10.2
F4 Reset Procedure
You may encounter an operating problem that blocks access to the Diagnostics Menu, or the
flow meter may operate erratically after exposure to a power transient or some other
traumatic event. These cases may require use of the F4-reset sequence to restore
operation.
The F4-Reset sequence operates on two levels:
●
Clear Active Memory
The first F4-Reset deletes all the data currently in Active Memory, but leaves Datalogger
data and all stored Site Setups intact. This is the most desirable method since all you have
to do to restore operation is reload a saved Site Setup.
●
Clear All Saved Memory
If the first sequence fails then you have to resort to the second level of the F4 sequence,
which allows you to clear ALL Saved Memory. Be aware that this erases all saved Site
Setups (including flow calibrated sites), Datalogger Data and user-defined pipe and sensor
tables. This will require you to completely re-install the system and repeat all desired default
settings, custom pipe tables, etc. The table below shows the sequence of the [F4] routine:
[Power On/Off + F4]⇒
[Clr Active Memory?]⇒
⇑ ⇓
[Clr Saved Data?]⇒
⇒No
⇒Yes
⇒No
⇒Yes
Clearing only Active Memory
1.
Turn off power (if it is currently on). Press <F4> and keep it pressed while you turn on
power. The prompt: [Clr Active Memory? No] appears at the top of the screen.
2.
Press <Right Arrow> to access F4 Reset option list. Press <Down Arrow> to switch the
option list to [Clr Active Memory? Yes]. Press <ENTER> to clear all Active Site Data (but
not saved Site Setups).
3.
To restore operation, press <MENU> to access the installation menu. Create a new site
setup or recall a stored site setup.
4.
Re-select any Meter Facilities menu items (e.g. RS-232 setup parameters).
Summary of Contents for 7ME3510
Page 2: ......
Page 10: ...Table of contents FUP1010 IP67 Portable 8 Operating Instructions 03 2012 A5E02951522 02 ...
Page 159: ...1010WP 7 21614 C ...
Page 160: ...1010WP 7 21614 C ...
Page 161: ...1010WDP 7 21614 C ...
Page 162: ...1010WDP 7 21614 C ...
Page 163: ......
Page 164: ......
Page 165: ......
Page 167: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Page 168: ......
Page 169: ...21614 C ...
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ...21614 C ...
Page 175: ...21614 C ...
Page 176: ...21614 C ...
Page 177: ...21614 C ...
Page 178: ...21614 C ...
Page 179: ...21614 C ...
Page 180: ...21614 C ...
Page 181: ......
Page 182: ......
Page 183: ...21614 C ...
Page 184: ...21614 C ...
Page 185: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Page 186: ...REFLECT MODE DIRECT MODE 21614 C ...
Page 187: ...1015BC 1 8 21614 C ...
Page 188: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Page 189: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Page 190: ......
Page 191: ......
Page 192: ......
Page 193: ......
Page 195: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Page 196: ......
Page 197: ......
Page 208: ...Glossary FUP1010 IP67 Portable 162 Operating Instructions 03 2012 A5E02951522 02 ...
Page 212: ...Index FUP1010 IP67 Portable 166 Operating Instructions 03 2012 A5E02951522 02 ...
Page 213: ......