Connecting
5.3 Zero Flow Adjust Menu
FUP1010 IP67 Portable
Operating Instructions, 03/2012, A5E02951522-02
63
To invoke Actual Zero:
1.
Access the [Zero Flow Adjust] option list by pressing <Right Arrow>.
2.
Press <ENTER>. A pop-up window prompts you to set the current flow rate (in selected
rate units) to equal zero (0.000).
Note
If a flow offset is desired (i.e., to test analog outputs) then press <Right Arrow> to enable
numeric entry.
3.
Press <ENTER> to start the Actual Zero process.
When you send the command, the flow meter analyzes the current flow rate for up to sixty
seconds, integrating (averaging) the data for the best zero correlation. During this time, the
menu prompt at the top of the display screen shows a timer that counts from zero to sixty.
You can allow zero averaging for the entire period, or cancel the process at any time by
pressing the <ENTER> key. This controls the amount of data the flow meter averages to
obtain a zero level.
Using ReversaMatic
If site conditions do not permit stopping the flow rate at the mounting location, and you do
not know the current flow rate, then you can use the ReversaMatic routine to establish the
zero flow level. You should perform the ReversaMatic procedure as quickly as possible to
ensure that the flow rate remains constant throughout the procedure.
To invoke ReversaMatic:
1.
To access the [Zero Flow Adjust] option list press <Right Arrow>.
2.
Move the cursor to [ReversaMatic]. Press <ENTER> to invoke the routine.
3.
The flow meter begins to measure the positive flow rate. "Positive" flow refers to flow
moving from upstream sensor location to the downstream sensor location. Note top
prompt line shows: Reversamatic Action
4.
Upon completion, the flow meter beeps and the display screen shows: Reverse Sensors /
Press <ENTER>
5.
Now remove then remount the upstream and downstream sensors in their reversed
positions. Mount the Up sensor (without removing its cable) in the Down sensor/cable
location. Mount Down sensor with its cable in the Up sensor/cable location. When
remounting the sensors, couple them to the pipe properly. Press <ENTER> (after re-
installing the sensors).
6.
The flow meter measures the negative flow rate briefly, then beeps and repeats the
prompt:
7.
Now remount the sensors for normal operation (in their original orientation). When
remounting sensors, couple them to the pipe properly. Press <ENTER> (after re-installing
the sensors).
Summary of Contents for 7ME3510
Page 2: ......
Page 10: ...Table of contents FUP1010 IP67 Portable 8 Operating Instructions 03 2012 A5E02951522 02 ...
Page 159: ...1010WP 7 21614 C ...
Page 160: ...1010WP 7 21614 C ...
Page 161: ...1010WDP 7 21614 C ...
Page 162: ...1010WDP 7 21614 C ...
Page 163: ......
Page 164: ......
Page 165: ......
Page 167: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Page 168: ......
Page 169: ...21614 C ...
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ...21614 C ...
Page 175: ...21614 C ...
Page 176: ...21614 C ...
Page 177: ...21614 C ...
Page 178: ...21614 C ...
Page 179: ...21614 C ...
Page 180: ...21614 C ...
Page 181: ......
Page 182: ......
Page 183: ...21614 C ...
Page 184: ...21614 C ...
Page 185: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Page 186: ...REFLECT MODE DIRECT MODE 21614 C ...
Page 187: ...1015BC 1 8 21614 C ...
Page 188: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Page 189: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Page 190: ......
Page 191: ......
Page 192: ......
Page 193: ......
Page 195: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Page 196: ......
Page 197: ......
Page 208: ...Glossary FUP1010 IP67 Portable 162 Operating Instructions 03 2012 A5E02951522 02 ...
Page 212: ...Index FUP1010 IP67 Portable 166 Operating Instructions 03 2012 A5E02951522 02 ...
Page 213: ......