Troubleshooting
10.3 Test Facilities Graph Screen
FUP1010 IP67 Portable
Operating Instructions, 03/2012, A5E02951522-02
133
Envelope Threshold Adjustment: (Hot Key 5 & 6)
Pressing the <=> key causes the graph to toggle between the default signal waveform
screen and the signal envelope screen (see example below). This envelope screen can aid
in the diagnosis of Tn errors caused by unusual receive waveform distortion. Signal
distortion is sometimes caused by poor sensor selection or poor pipe wall conditions, which
may result in an incorrectly measured fluid sound velocity. To improve the automatic
measurement of Tn, the envelope threshold limit can be adjusted to exclude portions of the
envelope, which may be causing the Tn detection problem.
)
96
>@
'
7Q
G7
6
61
+L6HW
①
④
Envelope Signal
Crossover Marker
②
⑤
HiSet Envelope Threshold
TN Marker
③
LoSet Envelope Threshold
⑥
Zero Baseline
Figure 10-4 Envelope Threshold Adjustment
1.
If it appears that the default placement of the Tn marker is incorrect or unstable, it can be
adjusted by pressing the <5> key on the keypad to invoke the [Hi Set #] command or by
pressing the <6> key to invoke the [Low Set #] command (while viewing the envelope
screen). A horizontal line representing the envelope threshold level will appear along with
a number indicating the percentage level. The High and Low thresholds can then be
moved either up or down on the envelope using the <+> or <-> keys. While viewing the
Tn marker position, adjust the thresholds so that they are well above the baseline "noise"
level but below the first major peak.
2.
To exit this mode, press the <0> key.
Summary of Contents for 7ME3510
Page 2: ......
Page 10: ...Table of contents FUP1010 IP67 Portable 8 Operating Instructions 03 2012 A5E02951522 02 ...
Page 159: ...1010WP 7 21614 C ...
Page 160: ...1010WP 7 21614 C ...
Page 161: ...1010WDP 7 21614 C ...
Page 162: ...1010WDP 7 21614 C ...
Page 163: ......
Page 164: ......
Page 165: ......
Page 167: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Page 168: ......
Page 169: ...21614 C ...
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ...21614 C ...
Page 175: ...21614 C ...
Page 176: ...21614 C ...
Page 177: ...21614 C ...
Page 178: ...21614 C ...
Page 179: ...21614 C ...
Page 180: ...21614 C ...
Page 181: ......
Page 182: ......
Page 183: ...21614 C ...
Page 184: ...21614 C ...
Page 185: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Page 186: ...REFLECT MODE DIRECT MODE 21614 C ...
Page 187: ...1015BC 1 8 21614 C ...
Page 188: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Page 189: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Page 190: ......
Page 191: ......
Page 192: ......
Page 193: ......
Page 195: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Page 196: ......
Page 197: ......
Page 208: ...Glossary FUP1010 IP67 Portable 162 Operating Instructions 03 2012 A5E02951522 02 ...
Page 212: ...Index FUP1010 IP67 Portable 166 Operating Instructions 03 2012 A5E02951522 02 ...
Page 213: ......