Connecting
5.2 Sensor Wiring
FUP1010 IP67 Portable
Operating Instructions, 03/2012, A5E02951522-02
29
Direct mount provides a shorter sonic beam path. This usually improves performance with
sonically attenuative liquids or pipe materials. We recommend using Direct mount for plastic
pipes. Compared to Direct mounting, Reflect mount requires almost double the amount of
mounting length. Therefore, Direct mount may be the only option if the availability of
mounting space is limited.
①
④
Upstream Sensor
Downstream Sensor
②
Pipe
⑤
Sonic path
③
Flow Direction
Figure 5-3
Direct Mount (Pipe shown from above in 12 o'clock position)
Mounting Supplies
The following items will be needed to mount the sensors (most are supplied):
●
Mounting Frames or Mounting tracks
●
Tape, chalk and a ruler or measuring tape
●
Spacer Bar
●
Mounting Guide (for Direct Mount)
●
Ultrasonic coupling compound and/or coupling pads
●
Sensors (matched set)
Selecting a location for the sensors
1.
Locate the sensors downstream from the center of the longest available straight run. A
location ten pipe diameters or greater downstream from the nearest bend will provide the
best flow profile conditions.
2.
Do not, if possible, install the sensors downstream from a throttling valve, a mixing tank,
the discharge of a positive displacement pump or any other equipment that could possibly
aerate the liquid. The best location will be as free as possible from flow disturbances,
vibration, sources of heat, noise, or radiated energy.
3.
Avoid mounting the sensors on a section of pipe with any external scale. Remove all
scale, rust, loose paint, etc., from the location.
4.
Do not mount the sensors on a surface aberration (pipe seam, etc.)
Summary of Contents for 7ME3510
Page 2: ......
Page 10: ...Table of contents FUP1010 IP67 Portable 8 Operating Instructions 03 2012 A5E02951522 02 ...
Page 159: ...1010WP 7 21614 C ...
Page 160: ...1010WP 7 21614 C ...
Page 161: ...1010WDP 7 21614 C ...
Page 162: ...1010WDP 7 21614 C ...
Page 163: ......
Page 164: ......
Page 165: ......
Page 167: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Page 168: ......
Page 169: ...21614 C ...
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ...21614 C ...
Page 175: ...21614 C ...
Page 176: ...21614 C ...
Page 177: ...21614 C ...
Page 178: ...21614 C ...
Page 179: ...21614 C ...
Page 180: ...21614 C ...
Page 181: ......
Page 182: ......
Page 183: ...21614 C ...
Page 184: ...21614 C ...
Page 185: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Page 186: ...REFLECT MODE DIRECT MODE 21614 C ...
Page 187: ...1015BC 1 8 21614 C ...
Page 188: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Page 189: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Page 190: ......
Page 191: ......
Page 192: ......
Page 193: ......
Page 195: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Page 196: ......
Page 197: ......
Page 208: ...Glossary FUP1010 IP67 Portable 162 Operating Instructions 03 2012 A5E02951522 02 ...
Page 212: ...Index FUP1010 IP67 Portable 166 Operating Instructions 03 2012 A5E02951522 02 ...
Page 213: ......