Chapter 4 Installation
Mobile DaRt Installation Manual
4-48
4.31
Checking Performance of Each Part
WARNING
Advance to the performance check after making sure of the connections
according to the supply voltage at the installation site.
4.31.1 Check tube current
Check tube current as following steps.
1.
Turn on the DIPswitch SW3-8 (adjustment mode) and SW2-5 (Measured
value display mode) of NEXSC board. Turn ON the key switch.
2.
Enter the tube current two-point adjustment mode. (See B.3)
3.
Check that tube current value displayed on panel is within the range below.
Adjust tube current when the tube current is not within the range.
1. 80kV, 50mA : 45 - 48 mA
2. 80kV, 160mA : 148 - 155mA