Appendix B Adjustment Mode
Mobile DaRt Installation Manual
B-viii
Tube current each point adjustment mode
Procedures of adjusting FVR value
1.
Set tube current each-point adjustment mode
Set the DIP switch NEXSC SW3-8 to ON and turn ON power. Then,
press the APR chest exposure key to enter the tube current each-point
adjustment mode. (LEDs of the chest exposure key, the regular position
exposure key, and the normal body exposure key blink.)
2.
Setting tube voltage and tube current
Fig. 5 Tube current each point adjustment mode – Exposure condition setting(In case of 160mA)
3.
Exposing X-RAY
Make an exposure by pressing the hand switch.
kV setting
mA setting