B.7 Display and Reset of Exposure Counter
Mobile DaRt Installation Manual
B-xv
B.7
Display and Reset of Exposure Counter
It is possible to display and reset the number of X-ray exposures that is
controlled by software.
Procedures of displaying the number of X-ray exposures:
1.
Setting exposure counter mode
Set DIP switch NEXSC SW3-8 ON and turn on power. Then, enter the
exposure counter display mode by pressing the Head exposure key of
APR. (The LED of the Head exposure key blinks.)
2.
Checking the cumulative number of exposures
The cumulative number of exposures appears on the set tube voltage
display and the display for the product of tube current and its flow time.
A max. number of cumulative number of exposures is 999999 times.
Over the counter’s max. number then, the cumulative number of
exposures is reset, and the display is turned to "0".
Fig. 13 Display of Exposure Counter (In case of 437 times)
Procedures of displaying the number of X-ray exposures:
1.
Resetting cumulative number of exposures
Press the program registration key for about 3 seconds. Then, the
cumulative number of exposures is reset, and the display is turned to
"0".
Summary of Contents for 503-61000-01
Page 2: ......
Page 6: ...No Text...
Page 14: ...Chapter 1 Introduction Mobile DaRt Installation Manual 1 8 No Text...
Page 22: ...Chapter 2 Outline Mobile DaRt Installation Manual 2 8 No Text...
Page 110: ...Chapter 6 Specifications Mobile DaRt Installation Manual 6 10 No Text...
Page 136: ...Appendix B Adjustment Mode Mobile DaRt Installation Manual B xxiv No Text...
Page 150: ...Appendix C Adjusting Methods Mobile DaRt Installation Manual C xiv No Text...
Page 154: ...Appendix D Error Message List Mobile DaRt Installation Manual D iv No Text...
Page 208: ...Appendix G Super Maintenance Function Mobile DaRt Installation Manual G iv No Text...