C.3
Adjusting XCONT-2002 board
Mobile DaRt Installation Manual
C-v
4.
Adjustment Max frequency of inverter.
Adjust VR2 [VFADJ(kV)] so that the frequency at CP29 [F] becomes 112
kHz (period: 8.9 to 9.0
μ
sec).
Fig. 1 Wave form at CP29[F]
5.
Rough adjustment of the inverter pulse width:
Performing READY and XRAY operations, adjust VR3 [PULSEADJ] so that the
pulse width at CP33 [A/D] becomes 7.5±0.1μsec. (Be cautious about the
polarity.)
Fig. 2 Wave form at CP33[A/D]
5V/div
7.5
±
0.1
μ
sec
5V/div
8.9-9.0
μ
sec
Summary of Contents for 503-61000-01