Shimadzu 503-61000-01 Installation Manual Download Page 10

  Chapter 1 Introduction 

 

 

  Mobile DaRt Installation Manual 

 

 

1-4 

 

Connect the power cord plug to a grounding 3P receptacle. 

 

Don’t splash this equipment with water because there is a risk of electric 

shock. And to clean the equipment, wipe the surface of the equipment 

with the clothes soaked in antiseptic solution. (Medical Alcohol) 

 

When operating the arm of the unit, make sure to place the unit on 

the floor whose inclination is 5 degrees or less. Otherwise, the unit 

may fall down if the arm position is not appropriate. 

 

Never drive the unit in a place where vibration will occur such as a 

road outdoor, a bumpy place or steep slope. Never stop the unit on a 

slope.  

 

Even if the bumper switch on the front of the unit contacts an 

obstacle while moving, obstacles are not detected if the force applied 

to the bumper switch is 2 kg or smaller. Watch out for obstacles in 

front while moving.   

 

This unit is driven by a built-in battery.   

Some wires are always exposed to 240 V of power voltage from the 

battery even when the key switch, switch for activating the 

emergency brake release function, or main circuit breaker is turned 

off. Watch out for an electric shock during operation. 

Wires that always have power voltage are red. 

 

The software (Operating system, OS) installed in this device runs 
under Windows XP. Do not change any of the OS settings. If the OS 
settings are changed, device operation after those changes is not 
guaranteed. 

CAUTION

WARNING

WARNING

WARNING

CAUTION

CAUTION

CAUTION

Summary of Contents for 503-61000-01

Page 1: ...Manual No M503 E320 Revision B Mobile X ray System MUX 100D INSTALLATION MANUAL Medical Systems Division This manual is for professional service engineers It bears no relation to the usual operation...

Page 2: ......

Page 3: ...Generator Communication S W 4 16 4 7 Setup of CXDI Generator Communication S W 4 19 4 8 Check of CXDI Generator Communication S W 4 27 4 9 Use permission setting of external storage medium 4 27 4 10...

Page 4: ...ecifications 6 1 Specifications 6 2 6 2 Dimensional Drawing of the Unit 6 7 6 3 X ray Reference Axis and Focal Spot Position 6 8 6 4 Exposure Condition 6 9 Appendix A NEXSC DIP Switch Settings A 1 Tab...

Page 5: ...iii E 3 CPU Backup battery Replacement E vii E 4 Collimator Lamp Replacement E viii E 5 Maintenance of rotary X ray tube section E ix E 6 Centering the Collimator and X ray Focus E xiv E 7 Adjustment...

Page 6: ...No Text...

Page 7: ...it may deliver its full performance and functions Besides thoroughly refer to the Operation Manual and Survice Manual of this unit Refer to Canon CXDI 50G Service Manual for installing CXDI 50G Chapte...

Page 8: ...ion terms used in the operation manual are defined as below States a direct danger that may cause death or serious injury if it is not avoided States an indirect or potential danger that may cause dea...

Page 9: ...When using X ray equipment incorrectly operator and patient may have unnecessary irradiation Be careful so that person other than patient does not stay in the examination room when X rays are exposed...

Page 10: ...the unit on a slope Even if the bumper switch on the front of the unit contacts an obstacle while moving obstacles are not detected if the force applied to the bumper switch is 2 kg or smaller Watch...

Page 11: ...age Be sure to follow the procedures described in this manual to start or stop this product Also do not turn the main power breaker OFF during normal operation of this device and especially during the...

Page 12: ...tware 1 Application i This software hereafter the software includes all software programs additional data and the Installation Guide used by this device 2 Rights i The Customer is allowed to use the s...

Page 13: ...s can be obtained Contact the responsible service shop of Shimadzu Corporation Document Document Number MUX 100D Operation Manual M503 E017 MUX 100D Installation Manual M503 E320 X ray tube assembly O...

Page 14: ...Chapter 1 Introduction Mobile DaRt Installation Manual 1 8 No Text...

Page 15: ...The name of each part of the unit is described in this Section 2 2 Name of Each Circuit Board and Layout of the Circuit Boards The name of each circuit board and layout of the main circuit boards ins...

Page 16: ...retch in the horizontal direction X ray tube unit The combination of X ray tube and collimator can turn around the axis of X ray focus This makes positioning and adjustment of the irradiation field an...

Page 17: ...trol panel Refer to Operation Manual M503 E017 Emergency brake release switch Used to release brake in the event of a problem with the running system P Main circuit breaker Turning off this breaker cu...

Page 18: ...2 2 Name of Each Circuit Board and Layout of the Circuit Boards Right side of the unit Left side of the unit X CONT 2002 NEXSC CPU MU Driver 2 Breaker NFB1 Under side is ON Bus Extender MUX Charge 04B...

Page 19: ...Chapter 2 Outline Mobile DaRt Installation Manual 2 5 Top of the unit battery Power 100D Power 100D Power 100D...

Page 20: ...ent without a package for transport and storage Ambient temperature 10 40 elative humidity 30 60 no dew condensation Atmospheric pressure 700 hPa 1060 hPa Power supply AC power System Single phase AC...

Page 21: ...only for the period of less than one month Relative humidity 10 60 Atmospheric pressure 700hPa 1060hPa When the unit is kept in storage for a long time until its installation it is recommended to cha...

Page 22: ...Chapter 2 Outline Mobile DaRt Installation Manual 2 8 No Text...

Page 23: ...ecessary preparations Chapter Contents 3 1 Tools Required for Installation Tools Required for Installation The list of tools required for installation is described in this Section 3 2 Unpacking The wo...

Page 24: ...bar L type spanners No 2 5 10 Nipper Plier Multimeter Insulation lock Grease Cutter Waste cloth Alligator clip Alcohol for cleaning Oscilloscope Hexagonal bolt M8 x 16 Only for exchanging X ray tube...

Page 25: ...000 04 1 Arm cover 503 47619 503 47620 1 1 Screws are attached on the tap hole on the arm and arm cover Spare parts 502 21566 1 Including 46 pieces of screw caps 5 pieces are the reserves Operation ma...

Page 26: ...n the main circuit breaker and key switch refer to Fig 3 1 and Fig 3 2 and move the unit to where X ray radiation is allowed and power outlet is provided The main circuit breaker is provided at the fr...

Page 27: ...breaker Otherwise there is a danger of receiving electric shock Name of Each Cover and Layout of the Covers There is a sequence of opening closing the covers as shown in Fig 3 3 For example to remove...

Page 28: ...tector changes color any damage incurred is outside the scope of the device s guarantee Do not overload the imaging unit with heavy objects This can damage or break the imaging unit s sensor Load limi...

Page 29: ...while operation CAUTION Do not pull the cable Do not pull the imaging unit when the cable is caught Doing the above may break the cable causing a fire or electric shock which may damage or break the u...

Page 30: ...est position Turn off the key switch and the main breaker 2 Unscrews the three screws as shown in Fig 3 7 3 Unscrews the three screws on the Top cover 4 Pull up the front part of the top cover Fig 3 8...

Page 31: ...er cover pull out forward and remove the cover Fig 3 9 Method of opening closing the front cover How to Open Close the Front Rear Cover 1 Unscrew the two screws as shown in Fig 3 10 a 2 Unscrew the fo...

Page 32: ...and remove it Fig 3 11 Screw cramp positions of the cable cover How to Open Close the Front Cover 1 Remove the front side cover and front rear cover 2 Remove the cable cover 3 Remove the cable guide a...

Page 33: ...ove the line switch for activating the emergency brake release function Remove frame front cover The switch separates by blowing down the lever Fig 3 13 How to remove the switch for activating the eme...

Page 34: ...xagon socket head cap screws at the two locations indicated in Fig 3 15 3 Remove the knob of the arm lock release lever by turning it counterclockwise 4 Pull out the top plate and the main cover of th...

Page 35: ...I Generator Communication S W 4 7 Setup of CXDI Generator Communication S W 4 8 Check of CXDI Generator Communication S W 4 9 Use permission setting of external storage medium 4 10 Installation of MLT...

Page 36: ...king and sending MUX 100D and 50G 4 26 Adjustment of Image quality 4 27 Installing the arm cover 4 28 Attaching of a knob for locking collimator rotation 4 29 Changing According to the Power Supply 4...

Page 37: ...ommunication S W Refer to 4 8 Installation of MLT M S W optional Refer to 4 10 Check of MLT M S W optional Refer to 4 11 Calibration Refer to 4 12 Selftest Refer to 4 13 Use permission setting of exte...

Page 38: ...fer to 4 28 Setting of XCONT Refer to 4 30 Changing According to the Power Supply Refer to 4 29 Installing options Refer to Chapter 5 Adjustment of Image quality Refer to 4 26 Start Assemble for the C...

Page 39: ...Loosen the bolts a and remove the screws b Loosen the bolts d and remove the bolts c The bolts b are Pass the sensor cable between the frame and the mount Fix the sensor cable so that the length betw...

Page 40: ...4 Connect the X ray I F cable attached to CXDI 50G Connect the X ray I F cable to the Power Box and terminals of XCONT board Fig 4 3 a b Fix the X ray I F cable just like fixing the CXDI 50G cable des...

Page 41: ...g 4 3 b Attach the cover of the Power Box 5 Fix the Power Box Detach four PADs on the Power Box Detach the plate that holds the Power Box Place the Power Box on the base plate and fix it with the reta...

Page 42: ...he power cable to the inverter and Power Box Refer to Fig 4 4 Roll the excessive cable 7 Connect the PC and Power Box with the LAN cable 8 Fix the rolled cable using a cord bracket Fig 4 5 Fixing the...

Page 43: ...Control S W Install CXDI Control S W CXDI RD according to the following procedures 1 Create CCR folder D CCR at PC s D drive 2 CD is insert the DVD ROM drive and the window is opened 3 Select and open...

Page 44: ...nstallation Manual 4 10 5 After it agrees to the use permission contract Yes is pressed Fig 4 8 6 Check the folder at the installation destination D CCR Next is pressed Fig 4 9 7 Begin the file copy i...

Page 45: ...is completed 4 5 Setup of CXDI 4 5 1 Input the serial number of CXDI 50G Input the serial number of CXDI 50G to communicate between CXDI 50G and DR system Please change the setting according to the pr...

Page 46: ...tion Mobile DaRt Installation Manual 4 12 3 It is input by the line of Enter item as 1 and presses the Enter key Fig 4 13 4 Setting Mode is displayed It is input as 0 and presses the Enter key Fig 4 1...

Page 47: ...isplayed It is input to the line of Enter item as 7 and presses the Enter key Fig 4 15 6 Capture Devices Configuration Table is displayed Enter The key is pressed and it advances it to the second line...

Page 48: ...ial number part of 0x written in the line on one step is input and the Enter key is pressed Fig 4 17 9 The Enter key is pressed until the menu of CCR SETUP MENU Esc to go back is displayed Check the d...

Page 49: ...layed 8 is input to Enter item and the Enter key is pressed Fig 4 19 11 It is displayed in the line of Enter Item Please push some keys to continue The Enter key is pressed Fig 4 20 12 Check the DR sy...

Page 50: ...e user setting screen Refer to CANON Inc CXDI manual s for details 4 6 Installation of CXDI Generator Communication S W The following two settings is needed 1 Installation of CXDI Generator Communicat...

Page 51: ...tallation condition input To the line of Select generator maker 3 Shimadzu To the line of Select anatomical code 2 Use To the line of Select photo timer 1 Not use To the line of Select through mode 1...

Page 52: ...ey is pressed Fig 4 23 5 It is confirmed that there is the following files in the folder D CCR cgbig3 dll cgbig3 ini cgstrtbl ini Fig 4 24 The installation of the CXDI Generator Communication S W is c...

Page 53: ...s necessary to change the setting of the CXDI by installing CXDI Generator Communication S W As follows the procedures After the S W is installed it is necessary to change the following settings of CX...

Page 54: ...Manual 4 20 3 SYSTEM INFORMATION screen is displayed ADMINISTRATOR SETUP is pressed Fig 4 27 4 ADMINISTRATOR SETUP screen is displayed MODULE is pressed Fig 4 28 5 HIS Module Information Setting scree...

Page 55: ...Generator Communication Module Information Setting screen is displayed Use Generator Communication Module is pressed Fig 4 30 7 Use Generator Communication Module becomes VALID CXDI GEN is selected F...

Page 56: ...Rt Installation Manual 4 22 9 It returns to Generator Communication Module Information Setting screen POST COND RECEIVE TIMEOUT TIME sec is selected Fig 4 33 10 A soft keyboard is displayed It is inpu...

Page 57: ...played Check the following settings of three points are correct OK is pressed Use Generator Communication Module VALID CXDI GEN cgbig3 dll POST COND RECEIVE TIMEOUT TIME sec 30 Fig 4 35 12 When you pr...

Page 58: ...PC is shutdown Afterwards the DR switch is pressed and CXDI 50G is started Fig 4 37 14 The exposure method is edited referring to CANON Inc CXDI manual SYSTEM is pressed on the screen where CXDI 50G...

Page 59: ...n is displayed The exposure method of changing the setting is selected example Skull AP Fig 4 40 17 EDIT PARA screen is displayed GEN Com Module is pressed Fig 4 41 18 A soft keyboard is displayed The...

Page 60: ...displayed OK is pressed Fig 4 43 20 The message of Save the changed parameter is displayed OK is pressed Fig 4 44 21 EDIT EXPOSURE MODE screen is displayed Other exposure methods are set repeating 16...

Page 61: ...e illuminated on the X Ray control panel The checking of CXDI Generator Communication S W is completed above 4 9 Use permission setting of external storage medium The acquired data can be preserved fr...

Page 62: ...on Manual 4 28 4 Setting Mode is displayed It is input as 0 and presses the Enter key Fig 4 47 5 The menu of CCR SETUP MENU Esc to go back is displayed It is input to the line of Enter item as 5 and p...

Page 63: ...numerical value is input by the following lines and the Enter key is pressed To the line of Disc store files 1 To the line of Disc store output F To the line of Window Center 2048 To the line of Windo...

Page 64: ...The Enter key is pressed until the menu of CCR SETUP MENU is displayed Check the menu was displayed the Esc key is pressed Fig 4 51 9 The menu of Welcome to CCR is displayed 8 is input to Enter item a...

Page 65: ...DaRt Installation Manual 4 31 10 It is displayed in the line of Enter Item Please push some keys to continue The Enter key is pressed Fig 4 53 11 CXDI is started SYSTEM button is pressed Fig 4 54 12 S...

Page 66: ...Manual 4 32 13 SYSTEM INFORMATION screen is displayed ADMINISTRATOR SETUP is pressed Fig 4 56 14 ADMINISTRATOR SETUP screen is displayed JPEG FOR EXT STORAGE is pressed Fig 4 57 15 JPEG FOR EXT STORA...

Page 67: ...dium Mobile DaRt Installation Manual 4 33 16 It is confirmed that Loseless button was displayed and presses EXIT Fig 4 59 17 OK is pressed Fig 4 60 18 OK is pressed Fig 4 61 The use permission setting...

Page 68: ...processing software MLT M installation 1 CD is insert the DVD ROM drive and the window is opened It is confirmed that there is the following files cdr2mltm dll Fig 4 62 2 cdr2mltm dll file is copied...

Page 69: ...f MLT M S W optional Checking the operation of MLT M installed by 4 10 Installation of MLT M S W optional is done according to the following procedures 1 SYSTEM is pressed on the screen where CXDI 50G...

Page 70: ...RMATION screen is displayed SYSTEM SETUP is pressed Fig 4 66 4 SYSTEM SETUP screen is displayed DISPLAYING PARAMETER CHANGE QA SCREEN DURING EXPOSURE is pressed Fig 4 67 5 DISPLAYING PARAMETER CHANGE...

Page 71: ...The check of MLT M is completed above 4 12 Calibration 4 12 1 Aging Perform aging of the unit while checking performance of X ray exposures Make exposures in step 1 to step 9 125 kV according to Table...

Page 72: ...is aprox 60kV and 2mAs at the exposure distance SID 1m The calibration error might occur because of differences the exposure distance SID and the X ray back scatter etc When the calibration error occu...

Page 73: ...l It is confirmed to pass all items 4 14 Installation of DMW PS optional As follows the procedures software DMW PS for the DICOM communication installation 1 CD insert the DVD ROM drive and the window...

Page 74: ...nction manual of CANON CXDI DMW PS 4 16 Check of DMW PS function optional 1 Connect to server and MUX 100D 2 Make list of patient data on server 3 Communicate server and CXDI for patient data Fig 4 73...

Page 75: ...Select the CXDI exposure mode button 3 Irradiation and save the image of the phantom 4 20 Check of REX value Check of REX value refer to CANON Inc CXDI manual 4 21 Check of DICOM storage and print fun...

Page 76: ...ayed normally on the touch monitor in about three seconds Exposure condition 50kV 2mAs SID 1m Exposure condition communicate PASSED FAILED Preceding clause exposure condition must be transmitted norma...

Page 77: ...ble cover and FPD box 3 CXDI 50G is detached from MUX 100D according to the procedure from 8 to 2 of the clauses 4 3 Assemble for the CXDI 50G 4 The covers are attached 4 25 Packing and sending MUX 10...

Page 78: ...the Cable Guide from the Middle Arm M5 bolt 2places Fig 4 74 a 3 Attach the Upper arm cover to the Arm M4 screw 4places Fig 4 74 b 4 Attach the Lower arm cover to the Arm M4 screw 6places Fig 4 74 c...

Page 79: ...Installation Manual 4 45 4 28 Attaching of a knob for locking collimator rotation Attach a knob for locking collimator rotation A bag is packed together with the knob and screws for adjustment of X r...

Page 80: ...included in the maintenance parts P N 502 21566 01 5 Turn on the main circuit breaker 6 Make sure the circuit breaker NFB1 is ON under side Be sure to change the connections according to the supply v...

Page 81: ...er switch setup SW name Silkscreen Setup Means ON Standard JP1 KC KC signal is outputted to the terminal X2 C C1 For High Voltage Trans D125PH C C1 JP2 C2 For High Voltage Trans D125PH C2 S Master res...

Page 82: ...4 31 1 Check tube current Check tube current as following steps 1 Turn on the DIPswitch SW3 8 adjustment mode and SW2 5 Measured value display mode of NEXSC board Turn ON the key switch 2 Enter the t...

Page 83: ...the Irradiation field adjusting knob completely open 2 Confirm the light field pushing the collimator lamp switch 3 Turn the Irradiation field adjusting knob completely close 4 Confirm the light fiel...

Page 84: ...o correct direction pushing the small movement switches 6 Confirm the unit stop pushing the both small movement switches at the same time NOTE Safety function works when the driving handle is pushed o...

Page 85: ...type NOTE Setting of battery type set factory shipping Changing this set code need not to be done when installation Set code Type of Battery Manufacturer 0 OLD PXL12072 Japan Storage Battery Manufact...

Page 86: ...ry type A wrong setting of code may causes the unit to malfunction and may cause electrical component to be damaged by a fire or fire occurs NOTE Standard Battery is recommended in order to be operate...

Page 87: ...are included in spare parts Side cover 4 places 8 places in total for left and right Collimator upper part 2 places FPD box 4 places Top cover 3 places Cable cover 4 places Front cover 4 places Front...

Page 88: ...ration manual 4 35 Installing an apron hanger Do when the user wants to change the position where the apron hanger is installed Installation procedure of apron hanger 1 Remove the screw two places of...

Page 89: ...ches on NEXSC Board The settings required to operate the options are described 5 2 Installing Remote Control Option How to install the remote control option is described 5 3 Installing Protective Scre...

Page 90: ...DIP switches according to the options No DIP switch settings are needed for the protective screen option large cassette box option or grid case option Fig 5 1 Arrangement of DIP switches DIP switch No...

Page 91: ...e the remote controller option effective Component of Remote Controller Option The Remote controller option P N 502 21343 03 is composed by the following parts Confirm whether all parts become complet...

Page 92: ...e remote controller in place of the removed cover of the arm joint section Use the resin top cover and socket head bolt that have been attached as they are 4 Connect and layout the cable extending fro...

Page 93: ...unit Fig 5 5 After finishing the preparations check the performance of the keys listed in Table 1 Transmit from the slash range shown in Figure 5 7 to the arm attaching part Fig 5 5 Remote Controller...

Page 94: ...Chapter 5 Installing the Options Mobile DaRt Installation Manual 5 6 Fig 5 7 Range where remote control can be operated 270 5m...

Page 95: ...all these parts are contained in the package Parts Name Quantity 1 Protective screen 1 2 Rail 1 3 Rail mount 2 Screw Flat head screw SUS M4x8 4 Bolt HEX SKT head bolt M4x10 2 Fig 5 8 Component of Prot...

Page 96: ...n Fix Arm lock release lever knob Put Rail mount on Arm lock joint section cover Fig 5 11 Mounted Rail mount 4 Put Top plate on Rail mount Screw Top plate on Rail mount Check the setting of Protective...

Page 97: ...are contained in the package Parts Name Quantity UL Quantity Except UL 1 Area Dosimeter 1 1 2 Area Dosimeter Adapter 1 1 3 JDAP2 Cable 1 1 4 Sub Weight 2 2 5 Rubber sheet 1 6 Cover 1 7 Cover Stay 2 B...

Page 98: ...ounting the dose area meter 3 Detach the arm cover Fig 5 15 Detach the arm cover 4 Connect the power cable of the dose area meter Relay the power cable under the arm The power cable is wired from the...

Page 99: ...llation Manual 5 11 Fig 5 17 Adjustment of vertical motion balance Adjustment procedure 5 The top cover is detached 6 The arm part is lowered most below and the weight part in the column is lifted to...

Page 100: ...pter 5 Installing the Options Mobile DaRt Installation Manual 5 12 7 Two supplementary weights are installed and the vertical motion balance of the arm part is adjusted Fig 5 19 Supplementary weight N...

Page 101: ...ile DaRt Installation Manual 6 1 Chapter 6 Specifications Chapter Contents 6 1 Specifications 6 2 Dimensional Drawing of the Unit 6 3 X ray Reference Axis and Focal Spot Position 6 4 Exposure Conditio...

Page 102: ...ngth 635 1200mm Support stand rotation range 270deg Tube rotation angle 180deg Tube rotation around tube axis Forward 90 deg Backward 20 deg Minimum distances between Focal Spot and Image Receptor 312...

Page 103: ...3 2ms 40 90kV 0 32 320mAs 91 100kV 0 32 280mAs 101 110kV 0 32 250mAs 111 120kV 0 32 220mAs The current time product that can be set with the X ray tube voltage 121 125kV 0 32 200mAs Range of the setti...

Page 104: ...16deg Nominal X ray tube Voltage Radiography 125kV X ray tube Assembly Max Heat 750kJ 1060kHU Max Continuous heat dissipation rate 120W 170HU s X ray Tube Max Anode heat dissipation rate 212kJ 300kHU...

Page 105: ...Eq Maximum irradiation time 1000msec Dimension 491 W x 477 D x 23 H mm with operation grip Grayscale 12bit 4 096 grayscale Option Grid Attenuation equivalent 2 3mmAl Eq For detail about CXDI please re...

Page 106: ...Chapter 6 Specifications Mobile DaRt Installation Manual 6 6 6 2 Dimensional Drawing of the Unit System dimensional drawing Fig 6 1 System dimensions...

Page 107: ...ocal spot position Fig 6 2 X ray reference axis and focal spot position Front and upper view Fig 6 3 X ray reference axis and focal spot position Side view View from this side Refer to Fig 9 3 4 M6 de...

Page 108: ...320 mAs is possible Since the tube current at 90kV and 100mAs is 100mA time is 1 0sec A setup of 125mA of tube current is possible to take an exposure with 0 8 50mAs The time at 90kV and 25mAs is 0 2...

Page 109: ...Since the tube current at 90kV and 100mAs is 80mA time is 1 25sec When a setup of 125mA of tube current is possible to take an exposure with tube voltage is lowered to 65kV mAs is 0 5 200mAs The time...

Page 110: ...Chapter 6 Specifications Mobile DaRt Installation Manual 6 10 No Text...

Page 111: ...Mobile DaRt Installation Manual A i Appendix A NEXSC DIP SWITCH SETTINGS DIP switch settings on the NEXSC board are described in this Chapter Chapter Contents A 1 Table of NEXSC DIP SWITCH SETTINGS A...

Page 112: ...isplay mode OFF Normal mode OFF SW2 6 RESERVED Normally OFF OFF SW2 7 RESERVED Normally OFF OFF SW2 8 Initial setting mode ON Initial setting mode OFF Normal mode OFF SW3 1 EEPROM Initialization ON In...

Page 113: ...of Tube Current B 4 Adjustment of the Handle B 5 Adjustment of the Battery Detection Circuit B 6 Setting Date and Time B 7 Display and Reset of Exposure Counter B 8 Display and Reset of Cumulated Mil...

Page 114: ...and then turn on power for the unit After just turning on power for the unit confirm TEC is displayed on the X ray control panel Fig 1 Initialize display on adjustment mode When the program register k...

Page 115: ...sed Battery voltage drop check mode Use this mode to check the condition of the battery when imaging is operated and load is applied Refer to B 12 Adjustment mode of the battery detection circuit Refe...

Page 116: ...Normally the tube current is adjusted in 1 tube current two point adjustment mode When necessary the tube current at each adjustment point is checked and adjusted in 2 Tube current each point adjustm...

Page 117: ...y for about 3 seconds NOTE Although it is possible to initialize FVR values with the initialization of NEXSC board the other important data are also initialized at the same time Others Others 1 When t...

Page 118: ...EDs of the head exposure key the regular position exposure key and the normal body exposure key blink 2 Set the tube voltage and tube current It is not necessary to set the tube voltage because it is...

Page 119: ...is adjusted to FVR 3 87 5 Registering FVR values Register the FVR value when the tube current has become an appropriate value To register FVR value press the program registration key Then a character...

Page 120: ...d turn ON power Then press the APR chest exposure key to enter the tube current each point adjustment mode LEDs of the chest exposure key the regular position exposure key and the normal body exposure...

Page 121: ...lue Allowable error within 15 5 Fig 6 Tube current each point adjustment mode FVR value adjusted this figure shows an example of the display the FVR is setting to FVR 3 87 5 Registering FVR values Reg...

Page 122: ...gauge As the adjustment position number is indicated on the tube voltage display apply the force in the indicated direction with a spring gauge referring to Table 1 and Fig 8 Tie a piece of string or...

Page 123: ...1 Forward left direction 1 kg 2 Forward left direction 2 kg 3 Backward left direction 1 kg 4 Backward left direction 2 kg 5 Forward right direction 1 kg 6 Forward right direction 2 kg 7 Backward right...

Page 124: ...urns to negative voltage mode by pressing the Foot key The detected battery voltage is displayed by six digits 2 Press mAs mAs key while measuring the actual voltage of batteries and Press the Program...

Page 125: ...e key of APR The LED of the hand exposure key blinks Fig 11 Setting Date In case of February 21 2005 2 Checking change of date Reveal the year month day display by pressing the remote control selectio...

Page 126: ...play by pressing the remote control selection key The LED of the key flashes Change hour with the set tube voltage change keys and 24 hour display Change minute with the and key for changing set value...

Page 127: ...ure key blinks 2 Checking the cumulative number of exposures The cumulative number of exposures appears on the set tube voltage display and the display for the product of tube current and its flow tim...

Page 128: ...The total cumulative mileage appears on the set tube voltage display and the relative cumulative mileage appears on the display for the product of tube current and its flow time The total cumulated m...

Page 129: ...B 8 Display and Reset of Cumulated Mileage Mobile DaRt Installation Manual B xvii Fig 14 Display of Relative Cumulated Mileage In case of 16km Fig 15 Display of Total Cumulated Mileage In case of 16km...

Page 130: ...rror Log mode LED of the Abdomen Exposure key and Obese key blink Fig 16 Display of Error Log In case of eighth error code was F68 2 Check error codes Error code is displayed on mAs indicator and log...

Page 131: ...y and APR waist exposure key LEDs of the waist exposure key and the obese body key blink 2 Check the cumulative number of charging times The number of charging times is displayed on the set tube volta...

Page 132: ...peed display mode Set the DIP switch NEXSC SW3 8 to ON and turn ON power Then enter the running speed display mode by pressing the obese body key and ARR foot exposure key LEDs of the foot exposure ke...

Page 133: ...to switch the running speed display from one to another of both wheel running speed km hour left wheel running speed km hour and right running speed by pressing the remote control choosing key Fig 19...

Page 134: ...00 msec Values are measured when 80 msec pasts after the exposure starts Perform exposure by pressing the hand switch After the exposure measured tube voltage and tube current are displayed on the exp...

Page 135: ...2 Battery Voltage Drop Check Mode Mobile DaRt Installation Manual B xxiii At present conditions of the unit and battery cannot be judged using the values provided in this mode Use the values as refere...

Page 136: ...Appendix B Adjustment Mode Mobile DaRt Installation Manual B xxiv No Text...

Page 137: ...uit boards or parts are replaced and readjustment is needed refer to these adjusting methods Chapter Contents C 1 Preperation C 2 Adjusting MUX CHARGE 04A board C 3 Adjusting XCONT 2002 board C 4 Adju...

Page 138: ...e options Refer to Appendix A 3 The setting of DIP SW1 1 on INVERTER UNIT board is matched with the OFF side Silk pattern is MUX 100 side 4 Move the arm out of parking position For displaying and expo...

Page 139: ...Hz Confirm that the waveform at CP14 50kHz shows approximately 50 kHz period 17 to 23 sec 3 Adjusting the inverter frequency of 120 Hz Adjust VR3 120Hz so that the waveform at CP16 120Hz shows 120 Hz...

Page 140: ...position It is unnecessary to make further adjustment of VR4 in normal case 3 Turn of f the Main breaker 4 Make sure the breaker NFB1 is OFF and connect the power plug 5 Turn on the DIPswitch SW3 8 ad...

Page 141: ...requency at CP29 F becomes 112 kHz period 8 9 to 9 0 sec Fig 1 Wave form at CP29 F 5 Rough adjustment of the inverter pulse width Performing READY and XRAY operations adjust VR3 PULSEADJ so that the p...

Page 142: ...EADJ of Xcont board Fig 3 Waveform between CP Ga CP Ea CP Gb CP Eb on INVERTER UNIT board 7 Turn OFF the power turn OFF NFB1 and remove the short circuit between CP 15V and CP5 VSIN Adjusting mAs circ...

Page 143: ...Measure the positive side of actual battery voltage the voltage between TER3 and TER5 on MUXPOWER board 3 Press mAs and mAs key so that the same voltage as the actual voltage is obtained When the same...

Page 144: ...eck pins on the XCONT board 5 Turn the main breaker ON Confirmation of ready operation 1 Turn ON the key switch 2 Confirm that the collimator lamp comes on by performing the collimator lamp lighting o...

Page 145: ...Fig 5 Waveform at the CP9 TKV In case of 80kV 160mA 3 If necessary press the Chest key get into the tube current each point adjustment mode and then check that measurement value is with error within 1...

Page 146: ...s on the XCONT board Procedure of confirming for battery charge 1 Turn the breaker NFB1 ON Power will be turned ON 2 Press Obese key to move to the adjustment mode Page 2 3 Press Leg key after about 1...

Page 147: ...he amount buzzer of loud sounds sounds at the time of X ray exposure 1 It goes into setting change mode Please refer to the change method of handling description 4 8 various setup If DIPSW 2 8 or 3 8...

Page 148: ...me control of the sheet panel which turns volume for volume control of a sheet panel and sounds at the time of key operation etc can be performed volume is large or small please adjust buzzer volume 1...

Page 149: ...C 8 Finish Mobile DaRt Installation Manual C xiii C 8 Finish Return the DIP SW on NEXSC board to its normal use condition...

Page 150: ...Appendix C Adjusting Methods Mobile DaRt Installation Manual C xiv No Text...

Page 151: ...Mobile DaRt Installation Manual D i Appendix D Error Message List Referring this error message list quickly resolve problems if any Chapter Contents D 1 Error Message List D...

Page 152: ...uring irradiation is more than 50 smaller than the set value F23 Measured X ray tube current during irradiation is smaller than 10mA F31 Measured filament current is more than 50 higher than the set v...

Page 153: ...y stop switch has being pressed F76 An illegal value was transmitted from CXDI EX Out of range APR No H1 With the set exposure condition the heat accumulated on the anode of the X ray tube is over tol...

Page 154: ...Appendix D Error Message List Mobile DaRt Installation Manual D iv No Text...

Page 155: ...on of the unit Chapter Contents E 1 Expendable Parts List E 2 Batteries Replacement E 3 CPU Backup Battery Replacement E 4 Collimator lamp Replacement E 5 Maintenance of rotary X ray tube section E 6...

Page 156: ...1 F12 F23 On CONTROL UNIT MUX chassis Fuse F1 On MUX CHARGE 04A board Fuse F6 On MUX POWER 99 board 0 5A 250V Slow blow 072 01664 26 Fuse F1 F2 On MUX POWER 99 board Fuse F21 F22 On CONTROL UNIT MUX c...

Page 157: ...in the battery unit Do not short circuit and contact between terminals The battery with a different kind type manufacture and residual quantity should not exist together The device will not work prope...

Page 158: ...d cap screws on both side Fig 2 a and remove the FPD box 4 Remove both of the side covers and the upper cover 5 Remove the Fix plates which are fixing the main frame and DR assy 6 Disconnect the cable...

Page 159: ...unit and open the cover by unscrewing the flat countersunk head screw shown in Fig 2 e and disconnect the wiring for the battery Take out all the batteries 12 Insert new batteries and connect the wire...

Page 160: ...5 Connect the wires to the terminal block in the reverse order from No 8 to No 1 of the numbers shown in Fig 2 d Be sure to connect them in order first from terminal of BAT 1 refer to Fig 2 d 17 Write...

Page 161: ...attery Replacement CPU Backup Battery replacement Parts Number of the CPU Backup Battery is 501 78646 Procedures of replacement for the CPU backup battery 1 Remove the right side cover 2 Turn on the k...

Page 162: ...rn Procedures of replacement for the collimator lamp 1 Turn the key switch off 2 Remove the lamp cover with the 4 screws 3 Remove the lamp screened cover with 2 screws 4 Change the lamp 5 Reinstall th...

Page 163: ...on of the X ray tube section is found to be abnormal or short of lubrication grease part and as shown in Fig 4 Fig 5 Fig 6 and Fig 7 for the detail of part A B and C Recommended grease Albania grease...

Page 164: ...d remove cover A 2 Unscrew locking screw b and make plunger c loose 3 Unscrew two M3 flat countersunk head screws d and remove cover C 4 Unscrew four M6 hexagon socket head cap screws e and remove bea...

Page 165: ...e removed from the holding frame on a table or the like Use a cushion under the tube 4 Unscrew locking nut i disc spring j and M5 hexagon socket head cap screw k and remove bearing B In this state bea...

Page 166: ...ocking nut o disc spring p detent washer q and thrust washer r 4 Pull out the bearing toward the front and remove it 5 After making sure that there is nothing abnormal with Part apply grease there 6 A...

Page 167: ...llowing pressure and torque Plunger in Part A After pushing it in to the limit with screwdriver turn it back 45 degrees to fix Locking nut in Part B Fig 6 i Adjust it so that the rotating torque becom...

Page 168: ...r illumination field center shifts from the X ray irradiation field center In this case align the center axis of the collimator blades with the axis of conical X ray beam by moving the collimator itse...

Page 169: ...each other using a dedicated device in factory assembling never move the lamp mounting board Fig 8 Adjust field center Fig 9 How to adjust center CAUTION Cassette Lamp field Solder Center of lamp fie...

Page 170: ...n 1 mm a portion called half shadow is physically occurs on the illumination field and due to the half shadow the illumination field becomes larger than the X ray irradiation field Discrepancy between...

Page 171: ...of the illumination field after determining the illumination field size in accordance with the exposing portion of the subject In this way discrepancy between the collimator illumination field center...

Page 172: ...MUX CHARGE 04A PCB for preventing lamp from getting excessive voltage during adjustment 2 Connect oscilloscope to CP17 IC M8 pin No 1 and GND on MUX CHARGE 04A PCB 3 Turn on the unit Turn on the key s...

Page 173: ...ltage r m s value will be 10 5 11 5V at lamp leads when the pulse width is adjusted to 3 8ms Make the pulse width wider when needed to increase illumination Make the pulse width narrower when needed t...

Page 174: ...h alcohol or the antiseptic was infiltrated Do cleaning and disinfections work certainly also to what has unevenness in the surface using a swab etc Do not splash antiseptic on this equipment Do not s...

Page 175: ...e set by infection and it may suffer from the infection which appears and is attacked with condition In connection with carrying out of the product in the case of recession from a medical institution...

Page 176: ...ed when MUX does X ray exposure Communications Specifications Communication protocol The communications protocol is described below 1 The master station requests establishment of a data link 1 The mas...

Page 177: ...CC horizontal parity Exclusive OR to ETX excluding STX Error Processing Parity check on each error detected data Horizontal parity checks on each message unit Physical Layer 1 Electrical specification...

Page 178: ...to permit establishment of the data link If the data link cannot be established the slave station sends NAK 3 If the slave station does not return ACK within the specified time RT1 in Table 2 the mas...

Page 179: ...link Master station Slave station ENQ NAK if receiving is evaluated as currently not possible Retransmission ACK if receiving is now evaluated as possible Establishing data link S T X Message E T X B...

Page 180: ...shing data link 2 0 sec 3 0 sec RT1 Number of ENQ retransmissions when establishing data link 3 times 3 times RT2 Number of retransmissions when sending message 3 times 3 times Maximum text length exc...

Page 181: ...to ETX excluding STX 1 A command contains the following three fields i Header The header is represented as a 2 byte code ii Message Expressed in JIS 8 bit code iii Terminator One byte field indicating...

Page 182: ...st response Radiographic condition instruction transmission B1 Implementation condition transmission 31 Implementation condition transmission 31 Radiographic condition instruction transmission B1 Impl...

Page 183: ...UX waits at least one second before transmitting ENQ for the 00 command 2 Circuit test response DR system MUX This is the response to the circuit test request in 1 above Header Message S T X 80 Test d...

Page 184: ...ion format Table 4 Massage details Item Bytes Description a 2B Manufacturer model code for connected instrument b 1B Radiography Fluoroscopy flag c 2B Tube Technique code d 8B Film related code disabl...

Page 185: ...as shown in the table below Table 5 Message details Item Description a Manufacturer model code for connected instrument 01 b Radiography Fluoroscopy flag Fixed as 0 c Tube code Technique code Tube co...

Page 186: ...re condition transmitted from CXDI to MUX is notified with an APR number Indicates the tube current used for radiography as 4 digits between 0010 and 160 mA Steps for MUX are 50 63 80 100 125 160mA Th...

Page 187: ...Leg Foot Arm Hand radiography direction 3 Front AP Side Lateral Oblique Body thickness 3 Thin Normal Obese 8 x 3 x 3 72types i Automatic exposure mechanism phototimer With this unit automatic exposure...

Page 188: ...to be set on CXDI when connecting with MUX With the unit exposure condition transmitted from CXDI to MUX is notified with an APR number 6 Exposure time Exposure time does not need to be set on CXDI wh...

Page 189: ...12 Chest Front AP Obese 82 3 2 13 Chest Side Lateral Thin 88 3 2 14 Chest Side Lateral Normal 91 3 6 15 Chest Side Lateral Obese 95 4 16 Chest Oblique Thin 87 3 2 17 Chest Oblique Normal 90 3 2 18 Che...

Page 190: ...n 57 2 50 Foot Side Lateral Normal 58 2 51 Foot Side Lateral Obese 61 2 52 Foot Oblique Thin 55 2 5 53 Foot Oblique Normal 58 2 5 54 Foot Oblique Obese 61 2 8 55 Arm Front AP Thin 58 2 8 56 Arm Front...

Page 191: ...eover after software is installed peculiar setting to customers of DICOM transmission and the print etc is done One recovery CD is attached by one PC It is not possible to restore it to other PC The u...

Page 192: ...S Features screen is displayed 5 First Boot Device is selected pushing key or key to the keyboard Fig 23 Advanced BIOS Features of selected First Boot Device The setting of First Boot Device is checke...

Page 193: ...e setting of First Boot Device is CDROM and pushes the Escape key to the keyboard It returns to the Setup utility screen Fig 24 The setting change screen of First Boot Device 7 Recovery CD is set in t...

Page 194: ...tion screen 2 This computer was started from a Ghost bootable CD is selected pushing key or key to keyboard It is checked to have selected 2 This computer was started from a Ghost bootable CD and push...

Page 195: ...continue in the screen It is repeated to push C key while pushing the Ctrl key to the keyboard until Terminate batch job Y N is displayed on the screen Fig 28 Terminate batch job Y N is displayed on...

Page 196: ...rom the keyboard is key push while pushing the Shift key to the keyboard Enter key to the keyboard is pushed Fig 30 display of c ghost inputting from the keyboard The processing of Ghost is restarted...

Page 197: ...n Manual E xlii i 15 Local is clicked Disk is clicked And From Image is clicked Fig 32 display of Image select 16 The drive C drive where recovery CD is set is clicked Fig 33 Display of Drive select 1...

Page 198: ...w is displayed OK is clicked Fig 35 Display of select local destination drive 19 The size of each partition when restoring it is checked OK is clicked Fig 36 Display of Destination Drive Details 20 If...

Page 199: ...gins Fig 38 Beginning of The restoration 22 If the screen below is displayed CD is replaced Fig 39 Display of Swap volume Recovery CD of the CD drive is replaced by opening the CD drive and the CD dri...

Page 200: ...ive Afterwards Reset Computer is clicked Fig 40 Display of Clone Complete 24 The restoration is finished Afterwards Install the software of CXDI like CXDI and the generator communication software etc...

Page 201: ...ppendix F Operation of High Voltage Cable Chapter Contents F 1 Spare parts with high voltage cable F 2 Installation procedure of the high voltage generator side plug F 3 Installation procedure of the...

Page 202: ...enerator side plug Contact pin P N 511 15061 11 Contact pin tool Silicon packing P N 532 23029 3mm thickness P N 532 23061 2mm thickness Please use silicon packing that the thickness is thin 2mm thick...

Page 203: ...the Silicone Rubber Jacket and Washer with the Silicone Oil 5 Fit the Silicone Washer over the Contact Pins as shown above 6 Insert the plug into the receptacle socket till the front key hits the rece...

Page 204: ...vided Flange onto the Ring Flange 5 Wet the Silicone Washer both side and the Rubber Gasket with the Silicon Oil 6 Fit the Rubber Gasket onto the Plug against the Ring Flange 7 Push the Silicone Washe...

Page 205: ...Mobile DaRt Installation Manual G i Appendix G Super Maintenance Function Chapter Contents G 1 Super Maintenance Function G...

Page 206: ...s to check the irradiation counter etc without opening the covers With this function X ray tube current can be easily checked because the unit is automatically brought into the measured value display...

Page 207: ...s brought into the mode that has been chosen 6 For the operations in each of the modes refer to 4 5 Initial Settings in Appendix B Adjusting modes How to finish the mode 7 Pressing the setting change...

Page 208: ...Appendix G Super Maintenance Function Mobile DaRt Installation Manual G iv No Text...

Reviews: