E.9 DR system - MUX communication
Mobile DaRt Installation Manual
E-xxx
iii
(h)
APR code (Protocol Number code)
Used to transmit APR numbers.
DR system send to Program Code for MUX APR memorized
exposure conditions.
Protocol Code
0001 to 0072
radiography area…… 8
Head, Chest, Abdomen, Waist, Leg, Foot, Arm, Hand
radiography direction……3
Front (AP), Side (Lateral), Oblique
Body thickness……3
Thin, Normal, Obese
8 x 3 x 3 = 72types
(i)
Automatic exposure mechanism (phototimer)
With this unit, automatic exposure mechanism (phototimer) is not
used.
Fixed as "0" because AEC system is not used this system. The film
sensitivity, film density, and phototimer detection field are ignored
the phototimer is not used.
(j) ID
code
Fixed as "00"
(k) Focus
code
Fixed as focus size-"1000"- large focus at this system.