E.5 Maintenance of rotary X-ray tube section
Mobile DaRt Installation Manual
E-xiii
Assemble the X-ray tube section so that the plunger in Part A, locking
nut in Part B, and the plunger and locking nut in Part C are fixed with
the following pressure and torque.
Plunger in Part A
After pushing it in to the limit with
screwdriver, turn it back 45 degrees to fix.
Locking nut in Part B
Fig. 6(i)
Adjust it so that the rotating torque
becomes 4.5
±
1.5 kg at the end of the
collimator handle.
Locking nut in Part C
Fig. 7(o)
After adjusting the plunger similarly as in
Part A, adjust it so that the rotating
torque becomes 4.5
±
1.5 kg at the end of
the collimator handle.
CAUTION