Appendix G Super-Maintenance Function
Mobile DaRt Installation Manual
G-ii
G.1
Super-Maintenance Function
This unit is provided with a function that can enter the initial setting and adjusting
mode (super-maintenance function) without setting DIP switches on NEXSC
board.
It is possible to adjust X-ray tube current, to make initial settings, to check the
irradiation counter, etc. without opening the covers. With this function X-ray tube
current can be easily checked because the unit is automatically brought into the
measured value display mode.
Super-maintenance function
1.
Initially enter the setting change mode.
2.
Press the head key
and collimator lamp key
in
succession within 3 seconds after display
has changed to display (setting of Power assist control)
3.
Display
appears for approx. one second,
and then display
(adjusting mode) appears.
Fig. G-1 Selecting the adjusting mode