User Manual SM6026
74
The FUNCtion:IMPedance command is used to set instrument functions. The FUNCtion:IMPedance?
query returns the current function parameters.
Command syntax: FUNCtion:IMPedance <function>
<function> can be one of the following items.
CPD
Set the function as Cp-D
LPRP
Set the function as Lp-Rp
CPQ
Set the function as Cp-Q
LSD
Set the function as Ls-D
CPG
Set the function as Cp-G
LSQ
Set the function as Ls-Q
CPRP Set the function as Cp-Rp
LSRS
Set the function as Ls-Rs
CSD
Set the function as Cs-D
RX
Set the function as R-X
CSQ
Set the function as Cs-Q
ZTD
Set the function as Z-θ◦
CSRS Set the function as Cs-Rs
ZTR
Set the function as Z-θr
LPQ
Set the function as Lp-Q
GB
Set the function as G-B
LPD
Set the funxtion as Lp-D
YTD
Set the function as Y-θ◦
LPG
Set the function as Lp-G
YTR
Set the function as Y-θr
For example: WrtCmd (“FUNC:IMP RX”) Set the function as R-X.
Query syntax: FUNCtion:IMPedance?
Return format: <function><NL^END>
The FUNCtion:IMPedance:RANGe command is used to set the range. The
FUNCtion:IMPedance:RANGe? query returns the current range.
Command syntax: FUNCtion:IMPedance:RANGe <value>
Where, <value> can be the impedance of the DUT or NR1, NR2 or NR3 data
format followed by OHM or KOHM.
For example: WrtCmd (“FUNC:IMP:RANG 1KOHM”) Set the range as 1kOHM.
Query syntax: FUNCtion:IMPedance:RANGe?
Return format: <value><NL^END>
Where, <value> can be
10
30
100
300
1000
3000
10000 30000 100000
The FUNCtion:IMPedance:RANGe:AUTO command is used to set the automaticrange selection status.
The FUNCtion:IMPedance:RANGe:AUTO? query returns the current range status.
Command syntax:
ON (1)
FUNCtion:IMPedance:RANGe:AUTO
OFF (0)
Where,
Character 1 (49) is equal to ON.
Character 0 (48) is equal to OFF.
For example: WrtCmd (“FUNC:IMP:RANG:AUTO ON”) Set the automatic range as ON.
Query syntax: FUNCtion:IMPedance:RANGe:AUTO?
Return format: <NR1><NL^END>
The FUNCtion:Source MONitor:VAC command is used to set the voltage monitor ON or OFF. The
FUNCTtion:Source MONitor:VAC? query returns the current voltage monitor status.