User Manual SM6026
31
MEAS
When the reference component is connected to the test terminal, you should press MEAS.
Then SM6026 will test the reference component and the test result will be automatically input
as the value of REF B.
4) Use MEAS or numeric keys to input the reference value of the secondary parameter. If the
reference values of primary and secondary parameters have been set in steps 2), you can skip
this step.
5) Move the cursor to the DEV A zone, the following soft keys will be displayed:
ΔABS
Δ%
OFF
6) Use above soft keys to se the deviation mode of the primary parameter.
7) Move the cursor to the DEV B zone, the following soft keys will be displayed.
ΔABS
Δ%
OFF
8) Use above soft keys to se the deviation mode of the secondary parameter.
3.6
<CORRECTION>
Press
[SETUP]
to select
CORRECTION
to enter into the
<CORRECTION>
Page.
Open, short and load correction on the
<CORRECTION>
page can be used to eliminate the distribution
capacitance, spurious impedance and other measurement errors. SM6026
provides two correction
modes: the first one is executing open and short correction on all frequency points through interpolation
method; the other one is executing open, short and load correction on the frequency point currently set.
The following measurement control parameters can be set on the
<Correction>
page.
Open correction
(OPEN)
Short correction
(SHORT)
Load correction
(LOAD)
Cable length selection
(CABLE)