User Manual SM6026
7
1.2 Technical Specifications
Measurement Parameters
:
L, C, R, lZl, lYl, X, B, G, D, Q, q, DCR, Vdc-Idc
LCR Test Parameter
:
L
0.00001 μH — 99.9999 kH
C
0.00001 pF — 9.99999 F
lZl, R, X, DCR
0.00001 W — 99.9999 MW
lYl, B, G
0.00001 μs — 99.9999 s
D
0.00001 — 9.99999
Q
0.00001 — 99999.9
q (Deg)
-179.999° — 179.999°
q (Rad)
-3.14159 — 3.14159
Δ%
-999.999% — 999.999%
Transformer Test Parameters :
DCR1 (primary, 2 terminal), DCR2 (secondary, 2 terminal), M (mutual
inductance),
N, 1/N, Phase, Lk (Leakage inductance), C (primary, secondary
capacitance), balance test
Basic Accuracy
:
LCR Parameters: 0.05%, Transformer Parameters: 0.1%
Test Frequency
:
20 Hz to 1 MHz
Resolution
:
10 mHz
Equivalent Circuit
:
Serial, Parallel
Ranging Mode
:
Auto, Hold
Trigger Mode
:
Internal, Manual, External, BUS
AC level
Voltage range of test signal:
10 mV — 2 Vrms
Min voltage resolution
:
100 μV, 3 digit input
Voltage Accuracy
:
ALC ON: 10% x set v 2mV, ALC OFF: 6% x set v 2mV
Current range of test signal:
100 μA — 20 mA
Min current resolution
:
1 μA, 3 digit input
Current Accuracy
:
ALC ON: 10% x set c 20 μA, ALC OFF: 6% x set v 20 μA
DC Bias Voltage Source
V /I range
:
0 V — ± 5V / 0 mA — ± 50 mA
Resolution
:
0.5 mV / 5 μA
Voltage accuracy
:
1% x set v 5 mV
ISO ON
:
For the bias test of inductance & transformer
Source Impedance (AC )
:
ISO ON: 100 W, ISO OFF:
30 W , 50 W & 100 W selectable
Source Impedance (DCR)
:
30 W , 50 W & 100 W selectable
Measuring time (> 10 kHz)
:
Fast : 13 ms/time, Med : 67 ms/time, Slow : 187 ms/time Plus the
refresh time of display character
Average Times
:
1 – 255