User Manual SM6026
71
Return format: <NR1><NL^END>
7.1.6
Output RESister subsystem commands
The Output RESister subsystem commands are mainly used to set the output internal resistor mode. The
Output RESister? query returns the current output internal resistance status.
Command syntax:
30
ORESister
50
100
10CC
For example: WrtCmd (“ORES 30”); Set the output internal resistance is 30 OHM.
Query syntax: ORESister?
Return format: <NR1><NL^END>
7.1.7
OUTPut subsystem commands
The OUTPut subsystem commands are mainly used to set the status of the 1A DC bias source and DC
isolation function.
Command tree:
The :HPOWer command is used to set the 1A DC bias source as ON or OFF. The :HPOWer? query
returns the current status of the 1A DC bias source. Before using this command, please ensure the
option component 1A DC bias source has been installed.
Command syntax:
ON
OUTPut:HPOWer
OFF
1
0
Where,
Character 1 (49) is equal to ON.
Character 0 (48) is equal to OFF.
Query syntax: OUTPut:HPOWer?
Return format: <INT><NL^END> or <OPT1><NL^END>, OPT means the 1A DC bias source is used, or
unused.
The :DC:ISOLation command is used to set the DC isolation function of the 100mA/10V DC bias source
as ON or OFF. The :DC:ISOLation? query returns the current status of the DC isolation function.
Command syntax:
ON
OUTPut:DC:ISOLation
OFF
1
0
Where,
OUTPut
: High POWer ON(1) (valid only when option is installed)
OFF(0)
:DC
:ISOLation ON
(
1
)
OFF
(
0
)