User Manual SM6026
88
<load1/2/3 A> is NR3 data format and the primary load correction data at frequency spot 1/2/3.
<load1/2/3 B> is NR3 data format and the secondary load correction data at frequency spot 1/2/3.
7.1.15 COMParator subsystem commands
The COMParataor subsystem commands are used to set the bin comparator function including ON/OFF
setting, Limit table setting.
Command tree:
The COMParator[STATe] command is used to set the comparator function as ON or OFF. The
COMParator[STATe]? query returns the current comparator state.
Command syntax:
ON
COMParator[:STATe]
OFF
1
0
Where,
1 (decimal 49) is equal to ON.
0 (decimal 48) is equal to OFF.
For example: WrtCmd (“COMP ON”)
Query syntax: COMParator[:STATe]?
Return format: <NR1><NL^END>.
The COMParator:MODE command is used to set the comparator mode. The
COMParator:MODE? query returns the current mode.
COMParator
[:STATe]
ON (1)
OFF (0)
:MODE Absolute TOLerance
Percent TOLerance
SEQuence
:TOLerance
:NOMinal <value>
:BIN<n> <low limit>,<high limit>
:SEQuence :BIN <BIN1 low limit>,<BIN1 high limit>,
<BIN2 high limit>,<BIN3 high limit>,
…… ,<BINn high limit>
:Secondary LIMit <low limit>,<high limit>
:Auxiliary BIN
ON (1)
OFF (0)
:SWAP
ON (1)
OFF (0)
:BIN
:CLEar
:COUNt
[:STATe]
:DATA?
:CLEar