User Manual SM6026
41
3.8.3
Sweep parameter setup
Move the cursor to the table to perform the setup of each sweep parameter:
FREQ (HZ)
,
LMT
,
HIGH
and
LOW
. Use numeric keys on the front panel to input the data of test frequency/level/bias and high/low
limit used to compare, as well as the selected primary/secondary use to compare. After setting, if some
inputs are unnecessary, you can execute the function of “Delete line” in the soft key zone to delete the
corresponding value.
In the bottom of the
LMT
zone, parameter A indicates that the primary parameters of the measurement
result are use to compare with the high and low limits of the table. Parameter B indicates that the
secondary parameters of the measurement result are used to compare with the high and low limits of the
table. “---“ means no compare. The soft key zone has corresponding items. Press the soft key
LMT A
,
“A” will be displayed in the
LMT
zone. When press the soft key
LMT B
, “B” will be displayed in the
LMT
zone. While press the soft key
OFF
, data in LMT zone and the corresponding high and low limits will be
cleared and be displayed as “---”.
4
[SYSTEM] and <FILE MANAGE>
4.1
<SYSTEM SETUP>
Press [System] and then SYSTEM SETUP to enter into the <SYSTEM SETUP> page shown as below.
On this page, most system setup items are displayed, such as instrument main function, beeper, PASS
beeper, FAIL beeper, language, PASS word, bus mode, GPIB address, TALK only, Bias SRC, baud rate,
data/time.
NOTE: Press LOG key to save the setup when finishing all the system items in this interface.
4.1.1
PASS BEEP
This zone is used to control and display the beep mode when the test result is qualified.
Operation steps for setting PASS BEEP
:
1) Move the cursor to PASS BEEP, the following soft keys will be displayed.
HIGH LONG
This soft key is used to select high and long beep.
HIGH SHORT
This soft key is used to select high and short beep.
LOW LONG
This soft key is used to select low and long beep.
TWO SHORT
This soft key is used to select two low and short beeps.