User Manual SM6026
40
10) Repeat step 9 until the high limit of BIN 9 is input. Then the cursor will automatically move to LOW of
2nd. Input the low limit of the secondary parameter.
11) The cursor will automatically move to HIGH of 2nd. Input the high limit of the secondary parameter.
3.8
<LIST SWEEP SETUP>
Press
[SETUP]
and then
LIST SWEEP
to enter into the
<LIST SWEEP SETUP>
page as shown below.
The list sweep function of SM6026 can perform auto sweep test for the test frequency, test level or bias
voltage of 10 points. On
<LIST SWEEP SETUP>
page, the following list sweep parameters can be set.
Sweep mode
(Mode)
Sweep parameter setup (frequency
[Hz]
, level
[V]
, level
[I],
bias
[V]
, bias —)
Sweep test point setup (sweep point)
Selection of limit parameter
(LMT)
High/low limit
(HIGH, LOW)
3.8.1
MODE
Mode menu is the same as the mode on
<List sweep display>
page.
3.8.2
Test parameter
Sweep parameters can be: frequency
[Hz]
, level
[V],
level
[I]
, bias
[V]
, bias
[I]
.
Operation steps for setting test parameter
:
1) Move the cursor the line following MODE; the following soft keys will be displayed.
FREQ [Hz]
LEVEL [V]
LEVEL [A]
BIAS [V]
BIAS [A]
2) Press one of above soft keys to select the list sweep parameter.