2-19
After Processing
Performing Ozone Cycle Manually
You can perform the ozone disinfection cycle manually by
pressing the "Manual Ozone Cycle" among the buttons on
the right side of the screen.
Select "Yes," and the ozone cycle operation progress
screen will appear.
If the window is open when "Yes" is selected, a message
appears and prompts you to close the window.
The screen shows the progress of operation. Once the
processing is complete, the screen closes.
The manual ozone cycle time is approx. 75 minutes.
To cancel, press the "Stop" button, and the ozone cycle
operation cancellation confirmation screen will appear.
For details, refer to the next item, "Cancelling Manual
Ozone Cycle."
If alcohol was used in the chamber or waste tank, dry
the chamber/tank completely. If any setting is made
when the chamber and/or waste tank is still wet with
alcohol, the instrument may detect an error or
malfunction.
Cancelling Manual Ozone Cycle
1. Pressing the "Stop" button on the ozone cycle operation
progress screen displays the ozone cycle operation
cancellation confirmation screen.
2. Select "Abort" and the ozone cycle cancellation screen
will appear.
The screen shows the progress of operation and once the
cancellation process is complete, the screen closes.
However, the ozone purge cycle cannot be cancelled.
Manual ozone cycle button
Summary of Contents for Tissue-Tek Polar DM
Page 1: ...Operating Manual Frozen Tissue Section Preparation Instrument Cryostat Microtome...
Page 15: ...Chapter 1 Basic Knowledge of Instrument...
Page 32: ...1 17 Standard Accessories Options 4 5 15...
Page 68: ...1 53...
Page 69: ...Chapter 2 Operating Method...
Page 108: ...2 39...
Page 109: ...Chapter 3 Maintenance Inspection...
Page 122: ...3 13...
Page 123: ...Chapter 4 Troubleshooting...