Daily Care and Inspection
How to Set the Draft Generator
The draft generator suctions air in the top part of the chamber to keep contaminated air from diffusing out of
the chamber.
1. Set the draft generator at the rear of the Polar® instrument.
2. Lift the rear cover of the Polar® instrument [1] and set the draft generator by shifting the pre-filter toward
the instrument (refer to p. 3-6) [2].
3. Plug in the connector leading from the Polar® instrument [3].
4. Turn on the power switch on the draft generator.
5. To use the draft generator, press the "Draft Generator" button on the [UTILITIES] Tab Screen and select
"ON."
If a connector cable must be connected to the Polar® instrument, contact the nearest authorized
Sakura distributor or representative.
Draft Generator Error Codes
If the draft generator experiences a clogged filter or fan abnormality, the LED lamp and suction level lamp
start blinking/turn to steady light to notify the situation. The lamp modes and meanings are listed below.
* If the particular situation you encountered is not listed above, contact the nearest
authorized Sakura distributor or representative.
Lamp colors and illumination patterns
Mode
LED lamp
Suction level lamp
Green (steady light) --
Normal operation mode
Green (blinking)
--
Standby mode
Green + red (lit
simultaneously)
--
The filter is clogged.
Replace the filers. (p. 3-6)
Red (blinking)
Blue (level 3, steady
light)
Fan revolutions are abnormal.
Turn off the power and then turn it
back on.
Red (blinking)
Blue (level 4, steady
light)
The fan has reached its life.
Contact the Sakura authorized
distributor.
Red (blinking)
Blue (level 5, steady
light)
This is a software error. Contact
the Sakura authorized distributor.
Suction force change switch
LED lamp
LED lamp
Power switch
Connector
LED lamp
(green or red)
Draft generator
Power switch
Suction level lamp
(1 to 5) (blue)
Suction force
change switch
LED lamp
(green or red)
Summary of Contents for Tissue-Tek Polar DM
Page 1: ...Operating Manual Frozen Tissue Section Preparation Instrument Cryostat Microtome...
Page 15: ...Chapter 1 Basic Knowledge of Instrument...
Page 32: ...1 17 Standard Accessories Options 4 5 15...
Page 68: ...1 53...
Page 69: ...Chapter 2 Operating Method...
Page 108: ...2 39...
Page 109: ...Chapter 3 Maintenance Inspection...
Page 122: ...3 13...
Page 123: ...Chapter 4 Troubleshooting...