![Sakura Tissue-Tek Polar DM Operating Manual Download Page 31](http://html1.mh-extra.com/html/sakura/tissue-tek-polar-dm/tissue-tek-polar-dm_operating-manual_3543386031.webp)
1-16
Specifications
General name
Frozen Tissue Section Preparation Instrument
Cryostat Microtome
Brand name
Tissue-Tek® Polar®
Model
POLAR-DM
POLAR-D
Instrument dimensions
785 mm (W) x 781 mm (D) x 1185 mm (H) (650 mm (W) without handwheel handle)
Chamber dimensions
500 mm (W) x 360 mm (D) x 280 mm (H)
(360 mm (D) excludes the evaporator)
Instrument mass
Approx. 176 kg
Approx. 174 kg
Rated power supply
100 VAC 50/60 Hz 15 A, 115 VAC 50/60 Hz 13 A, 230 VAC 50/60 Hz 6.5 A
Temperature range
Specimen head (specimen holder): -10
°
C to -50
°
C (room temperature: 25
°
C)
Chamber: -10
°
C to -35
°
C (room temperature: 15
°
C)
Cryobar temperature
-40
°
C or below (room temperature: 25
°
C)
Cryobar specimen capacity
15 (round type)
Cryo+ temperature
-55
°
C or below (room temperature: 25
°
C)
Sectioning thickness setting
range
1 to 99 μm
Maximum specimen size
∅
48.5 mm
Specimen orientation
XY-axes 8
°
/Z-axis 360
°
Blade angle adjustment range
4
°
to 14
°
Horizontal travel of blade holder
±
20 mm from the specimen mounting base
Sectioning auxiliaries
Anti-Roll Plate, Anti-Roll Rake
Defrost cycle
Hot gas/heater
Anti-infection functions
Ozone disinfection cycle: Approx. 75 minutes (including 15 minutes of ozone decomposition)
Ozone concentration: Approx. 1000 ppm
Ozone decomposition: Manganese catalyst
Draft generator
Optional (HEPA filter capable of removing at least 99.99% of 0.3
µ
m or larger particles)
Vacuum function
Standard equipment (pre-filter capacity: 164 cm
3
)
Sectioning method
Manual/powered
Manual
Motorized sectioning speed
12 to 200 mm/sec
−
Display
6.2" VGA TFT color LCD display (640 x 240 dots)
External interfaces
USB, LAN
Emergency Stop Switch
Standard equipment
Facility requirement
Single-phase 100/115 VAC
±
10%, 50/60 Hz, 15 A or more, with grounding terminal
Single-phase 230 VAC
±
15%, 50/60 Hz, 7 A or more, with grounding terminal
Operating environment
Operating temperature: 10 to 35
°
C
Operating humidity: 30 to 85% RH (non-condensing)
Protection against electric shock Class I
Pollution Degree 2
Installation category
II
* The equipment specifications are subject to change without notice.
Summary of Contents for Tissue-Tek Polar DM
Page 1: ...Operating Manual Frozen Tissue Section Preparation Instrument Cryostat Microtome...
Page 15: ...Chapter 1 Basic Knowledge of Instrument...
Page 32: ...1 17 Standard Accessories Options 4 5 15...
Page 68: ...1 53...
Page 69: ...Chapter 2 Operating Method...
Page 108: ...2 39...
Page 109: ...Chapter 3 Maintenance Inspection...
Page 122: ...3 13...
Page 123: ...Chapter 4 Troubleshooting...