![Sakura Tissue-Tek Polar DM Operating Manual Download Page 47](http://html1.mh-extra.com/html/sakura/tissue-tek-polar-dm/tissue-tek-polar-dm_operating-manual_3543386047.webp)
1-32
Registering/Editing in the Temperature Setting FREE Mode
When the temperature setting FREE mode is selected on the [SECTION] Tab Screen, you can change the
chamber temperature and specimen head temperature freely.
How to Set the Chamber Temperature
1. Press the "Chamber Temperature Setting" button on the
[SECTION] Tab Screen.
2. The chamber temperature setting screen appears.
The screen shows the measured value (current temperature)
and setting value (temperature to be set).
3. Press the "
"/"
" buttons among the "Value Change" buttons
to change the value of temperature.
The settable range is -10 to -35
°
C. A desired value can be set
as a multiple of 1
°
C.
4. Pressing the “Enter” button saves the setting, after which the
display returns to the [SECTION] Tab Screen.
If no operation has been performed for 5 minutes, the setting is
not saved and the screen closes automatically.
How to Set the Specimen Head Temperature
1. Press the "Specimen Head Temperature Setting" button on the
[SECTION] Tab Screen.
2. The specimen head temperature setting screen appears.
The screen shows the measured value (current temperature)
and setting value (temperature to be set).
3. Press the "
"/"
" buttons among the "Value Change" buttons to
change the value of temperature.
The settable range is -10 to -50
°
C. A desired value can be set as
a multiple of 1
°
C.
4. Pressing the “Enter” button saves the setting, after which the
display returns to the [SECTION] Tab Screen.
If no operation has been performed for 5 minutes, the setting is
not saved and the screen closes automatically.
If the instrument is not used for a specified time, it will enter the
standby mode (refer to p. 2-3) and the specimen head temperature
will remain at -20
°
C regardless of the setpoint temperature. Touch
the screen to cancel the standby mode, and the temperature will
be adjusted to the setpoint temperature.
Chamber temperature
setting button
Specimen head temperature
setting button
Setting value (setpoint
temperature)
Value change
button (up/down)
Enter button
Measured value (current chamber temperature)
Measured value (current specimen head temperature)
Setting value (setpoint
temperature)
Value change
button (up/down)
Enter button
Summary of Contents for Tissue-Tek Polar DM
Page 1: ...Operating Manual Frozen Tissue Section Preparation Instrument Cryostat Microtome...
Page 15: ...Chapter 1 Basic Knowledge of Instrument...
Page 32: ...1 17 Standard Accessories Options 4 5 15...
Page 68: ...1 53...
Page 69: ...Chapter 2 Operating Method...
Page 108: ...2 39...
Page 109: ...Chapter 3 Maintenance Inspection...
Page 122: ...3 13...
Page 123: ...Chapter 4 Troubleshooting...