![Sakura Tissue-Tek Polar DM Operating Manual Download Page 40](http://html1.mh-extra.com/html/sakura/tissue-tek-polar-dm/tissue-tek-polar-dm_operating-manual_3543386040.webp)
1-25
Motorized Sectioning Tab Screen (DM only)
PAUSE Mode Screen
In the PAUSE mode, final shaving is performed to the set section thickness, just like in the FINE mode, but
you can pause the sectioning every time the preset sectioning end position is reached, before the section is
cut off. Refer to p. 1-35 for information on how to change the stop position.
The section thickness and sectioning end position can be changed in this mode.
The chamber temperature and specimen head temperature in the temperature setting mode selected on
the SECTION Tab become effective.
To change these temperatures, change them in the FREE mode or Temperature Setting 1/2 mode.
•
Motorized sectioning mode selection button
Press this button to select and switch to the "TRIM," "FINE" or "PAUSE" mode.
•
Section thickness setting button
Press this button to change the thickness to be used when sectioning. Refer to p. 1-30 for details.
•
Sectioning end position adjustment button
You can adjust the position at which to stop the sectioning before the section is cut off. Refer to p. 1-35
for details.
Motorized sectioning
mode selection button
* The icon corresponding
to the currently selected
mode is displayed.
Sectioning end position adjustment button
Section thickness setting button
Currently set sectioning end position
Summary of Contents for Tissue-Tek Polar DM
Page 1: ...Operating Manual Frozen Tissue Section Preparation Instrument Cryostat Microtome...
Page 15: ...Chapter 1 Basic Knowledge of Instrument...
Page 32: ...1 17 Standard Accessories Options 4 5 15...
Page 68: ...1 53...
Page 69: ...Chapter 2 Operating Method...
Page 108: ...2 39...
Page 109: ...Chapter 3 Maintenance Inspection...
Page 122: ...3 13...
Page 123: ...Chapter 4 Troubleshooting...