![RS PRO 917-6298 Instruction Manual Download Page 205](http://html.mh-extra.com/html/rs-pro/917-6298/917-6298_instruction-manual_1496054205.webp)
IIT-2000/2010/5000/5010 Instruction Manual/English
05/08/2015 Version No. 001
203
LOW SET:
0.010mA to 1.099mA
00.10mA to 10.99mA
001.0mA to 039.9mA(IIT-5010/5000 is 099.9mA)
Window Comparator Method Yes
ARC DETECT
Yes
Rise-time Control Function
Yes
RAMP (Ramp Time)
0.1~999.9S
TIMER (Test Time)
OFF
1
, 0.5S~999.9S
GND
ON/OFF
1
The timer can only be turned off under special MANU mode (MANU=***-000).
DC Withstanding Voltage
Output Voltage Range
0.100kV ~ 6.000kV
Output Voltage Resolution
2V
Output Voltage Accuracy
(1% of s5V) with no load
Maximum Rated Load
50W (5kV/10mA)[IIT-2000/2010]
100W (5kV/20mA)[IIT-5010/5000]
Maximum Rated Current
10mA [IIT-2000/2010, 20mA [IIT-5010/5000]
0.001mA ~ 2mA (0.100
kV≤V≤0.5kV)
0.001mA ~
10mA (0.5kV<V≤6kV)[IIT-2000/2010]
0.0
01mA ~ 20mA (0.5kV<V≤6kV)[IIT-5000/5010]
Voltmeter Accuracy
(1% of 5V)
Voltage Regulation
1% +5V
[Maximum rated load
→ no load]
Current Measurement
Range
0.001mA~010.0mA [IIT-2000/2010]
0.001mA~020.0mA [IIT-5010/5000]