produced. All digits outside the HI SET range are ignored and will
not be used.
For example:
HI SET value: 12.34
LO SET value1: 0.005
error
LO SET value2: 0.053
no error
In the example above LO SET value1 will produce an error as all
digits are outside the range of HI SET. LO SET value2 will not
produce an error, but will return 0.05, not 0.053.
Syntax
Query Syntax
MANU:ACW:CLOSet<NR2>
MANU:ACW:CLOSet?
Parameter/
Return
parameter
<NR2> 0.000 ~ 041.9 (IIT-2000/2010)
0.000 ~ 109.9 (IIT-5010/5000)
Example
MANU:ACW:CLOS 20.0
Sets the ACW LO SET current to 20 mA.
MANU:ACW:TTIMe
Set
Query
Description
Sets or returns the ACW test time in seconds. The test must first
be in ACW mode before this command can be used.
Note: A “TIME ERR” will result if the Ramp Time + Test Time is ≥