IIT-2000/2010/5000/5010 Instruction Manual/English
05/08/2015 Version No. 001
119
4. Use the UP/DOWN arrow keys to choose a menu
item: Start Ctrl, Double Action, Key Lock or
INTERLOCK.
5. Use the scroll wheel to select setting for the chosen
menu item.
Start Ctrl
Double Action
Key Lock
INTERLOCK
FRONT PANEL, REMOTE CONNECT,
SIGNAL IO
ON, OFF
ON, OFF
ON, OFF
6. Press EDIT/SAVE to save the settings and exit to
VIEW status.
EDIT/SAVE
Note
The Double Action setting is ignored when the IIT-2000/
2010/5000/5010 is being controlled remotely using the USB,
RS232 or GPIB interfaces.
Note
If a test is started with INTERLOCK ON, but the interlock
signal I/O pins are not shorted (either with the included
interlock key or manually), the INTERLOCK OPEN message
will be displayed, preventing the test from starting.