IIT-2000/2010/5000/5010 Instruction Manual/English
200
05/08/2015 Version No. 001
R ERR
For IR tests. The voltage is too high or resistance=0Ω.
Check to see whether the DUT or test lead is shorting.
For GB tests. The resistance is too high.
I<SET
For GB tests. Current too low. Indicates that the SOURCE
L or SOURCE H test lead is open or poorly connected.
Test the test lead connection with the DUT to confirm.
I>SET
For GB tests. Current is too high.
R=0
For GB tests. Resistance = 0. This error indicates that
there is an error with the measured resistance (0Ω).
Perform the zeroing function again.
GBV OVER
GBV > 5.4V