IIT-2000/2010/5000/5010 Instruction Manual/English
05/08/2015 Version No. 001
165
Example
MANU:DCW:CLOS 2.00
Sets the DCW LO SET current to 2mA.
MANU:DCW:TTIMe
Set
Query
Description
Sets or returns the DCW test time in seconds. The test must first
be in DCW mode before this command can be used.
In special MANU mode, the TIMER can be turned off.
Syntax
Query Syntax
MANU:DCW:TTIMe {<NR2>|OFF}
MANU:DCW:TTIMe?
Parameter
<NR2>
OFF
0.5 ~ 999.9 seconds
TIMER OFF (special MANU mode).
Return
parameter
<NR2>
TIME OFF
0.5 ~ 999.9 seconds
TIMER is OFF (special MANU mode).
Example
MANU:DCW:TTIM 1
Sets the DCW test time to 1 second.
MANU:DCW:REF
Set
Query
Description
Sets or returns the DCW reference value in mA. The test must
first be in DCW mode before this command can be used.
The reference value must be less than the HI SET value.
The reference value uses the same range as the HI SET value.