IIT-2000/2010/5000/5010 Instruction Manual/English
172
05/08/2015 Version No. 001
Note
If the (GB current x HI SET resistance) > 5.4V, then an error will
be generated (“GBV > 5.4V”).
MANU:GB:RLOSet
Set
Query
Description
Sets or returns the GB LO SET resistance value in mΩ. The LO
SET value must be less than the HI SET value. The test must
first be in GB mode before this command can be used.
Syntax
Query Syntax
MANU:GB:RLOSet<NR2>
MANU:IR:RLOSet?
Parameter/
Return
parameter
<NR2> 0.000 ~ 649.9
Example
MANU:GB:RLOS 50
Sets the GB LO SET resistance to 50mΩ.
MANU:GB:TTIMe
Set
Query
Description
Sets or returns the GB test time in seconds. The test must first be
in GB mode before this command can be used.
Syntax
Query Syntax
MANU:GB:TTIMe <NR2>
MANU:GB:TTIMe?