IIT-2000/2010/5000/5010 Instruction Manual/English
156
05/08/2015 Version No. 001
HI SET
1.000mA
1.000mA
∞MΩ
100.0mΩ
LO SET
0.000mA
0.000mA
0001MΩ
000.0mΩ
I or V
V=0.100kV
V=0.100kV
V=0.050kV
03.00A
TIMER
001.0S
001.0S
001.0S
001.0S
RAMP
↗
000.1S
000.1S
000.1S
X
Example
MANU:INITial
Loads the initial settings for the selected MANU number.
MANU:RTIMe
Set
Query
Description
Sets or returns the Ramp Time for the test in seconds.
Note: A “TIME ERR” will result if the Ramp Time + Test Time is ≥
240 seconds when the HI SET limit is over 30mA (IIT-2000/2010)
or over 80mA (IIT-5010/5000). This applies to the ACW function
only.
Syntax
Query Syntax
MANU:RTIMe <NR2>
MANU:RTIMe?
Parameter/
Return
parameter
<NR2> 0.1~999.9 seconds
Example
MANU:RTIM 0.5
Sets the ramp time to half a second.