IIT-2000/2010/5000/5010 Instruction Manual/English
05/08/2015 Version No. 001
171
MANU:GB:CURRent
Set
Query
Description
Sets or returns the GB current in A. The test must first be in GB
mode before this command can be used.
Syntax
Query Syntax
MANU:GB:CURRent <NR2>
MANU:GB:CURRent?
Parameter/
Return
parameter
<NR2> 3.00~33.00 (IIT-2010)
3.00~33.00 (IIT-5010)
Example
MANU:GB:CURR 3.00
Sets the GB current to 3.00A.
MANU:GB:RHISet
Set
Query
Description
Sets or returns the GB HI SET resistance value in mΩ. The test
must first be in GB mode before this command can be used.
Syntax
Query Syntax
MANU:GB:RHISet <NR2>
MANU:GB:RHISet?
Parameter/
Return
parameter
<NR2> 000.1 ~ 650.0
Example
MANU:GB:RHIS 100.0
Sets the HI SET value to 100mΩ.