Parameter/
Return
parameter
<NR2> 0.5 ~ 999.9 seconds
Example
MANU:GB:TTIM 1
Sets the GB test time to 1 second.
MANU:GB:FREQuency
Set
Query
Description
Sets or returns the GB test frequency in Hz. The test must first be
in GB mode before this command can be used.
Syntax
Query Syntax
MANU:GB:FREQuency {50|60}
MANU:GB:FREQuency?
Parameter/
Return
parameter
50
50 Hz
60
60 Hz
Example
MANU:GB:FREQ 50
Sets the GB test frequency to 50Hz.
MANU:GB:REF
Set
Query
Description
Sets or returns the GB reference value in mΩ. The test must first
be in GB mode before this command can be used.
Syntax
Query Syntax
MANU:GB:REF <NR2>
MANU:GB:REF?