IIT-2000/2010/5000/5010 Instruction Manual/English
05/08/2015 Version No. 001
157
MANU:EDIT:MODE
Set
Query
Description
Sets or returns the mode (ACW, DCW, IR, GB) of the selected
manual test.
Syntax
Query Syntax
MANU:EDIT:MODE {ACW|DCW|IR|GB}
MANU:EDIT:MODE?
Parameter/
Return
parameter
ACW
AC Withstand mode
DCW
DC Withstand mode
IR
Insulation Resistance mode
GB
Ground Bond mode
Example
MANU:EDIT:MODE ACW
Sets the mode to ACW.
MANU:ACW:VOLTage
Set
Query
Description
Sets or returns the ACW voltage in kV. The test must first be in
ACW mode before this command can be used.
Syntax
Query Syntax
MANU:ACW:VOLTage <NR2>
MANU:ACW:VOLTage?
Parameter/
Return
parameter
<NR2> 0.100 ~ 5.000 (kV)