Instrument Function
R&S
®
SMBV100A
342
Operating Manual 1407.6062.32 ─ 08
Note:
The marker trace in the waveform file remains unchanged in any case. It is not
overwritten, even if something else than Unchanged has been selected.
"Unchanged"
The marker signal remains unchanged as defined in the waveform file.
"Restart"
A marker signal is generated at the start of each ARB sequence.
This marker mode is not enabled for muti segment waveforms.
"Pulse"
A regular marker signal is generated. The pulse frequency is defined
by entering a divider. The frequency is derived by dividing the sample
rate by the divider. The input box for the divider opens when "Pulse" is
selected, and the resulting pulse frequency is displayed below it.
The precision of the frequency setting depends on the sampling rate.
The maximum pulse frequency is equal to half of the sampling rate.
SCPI command:
"Pattern "
A marker signal that is defined by a bit pattern is generated. The pattern
has a maximum length of 64 bits and is defined in an input field which
opens when "pattern" is selected.
The pattern bits switch the marker signal to high and low state.
SCPI command:
"ON/OFF
Period"
A regular marker signal that is defined by an ON/OFF ratio is generated.
A period lasts one ON and OFF cycle.
The "ON Time" and "OFF Time" are each expressed as a number of
samples and are set in an input field which opens when ON/OFF ratio
is selected.
SCPI command:
SCPI command:
Marker Delay
The delay of the signals on the MARKER outputs is set in the"Marker Delay" section.
Marker x Delay
Enters the delay between the marker signal at the marker outputs and the start of the
frame or slot.